Troubleshooting
Continued from page before
Continued on next page
Indicator
Possible reason
Solution
By
Shorting plug can
not be removed.
The connector is
covered with ice
(frozen moisture or
nitrogen ice).
Carefully remove the ice with
warm helium gas. Use the
dipstick or the precooling tube
as tubing for the warm helium
gas to remove small ice spots
from the connector.
Bruker Service
The magnet system
quenches.
Loss of
superconductivity.
See chapter
Bruker Service
The helium level was
too low for energiz-
ing, cycling,
shimming,
de-energizing or
sweeping.
See chapter
Bruker Service
The Power Supply is
defective. The main
current is oscillating.
Replace the Power Supply.
Bruker Service
The main coil switch
can not be closed on
field.
The helium level is
too low for energiz-
ing. The main coil
switch is not covered
with liquid helium.
Never try to energize the
magnet with less than the
"minimum level during ener-
gizing" in the helium vessel.
Bruker Service
The Power Supply is
defective. The main
current is oscillating.
Replace the Power Supply.
Bruker Service
Shim current can not
be set correctly.
The control cable is
not connected
correctly to the
current lead or to the
Power Supply.
Connect the control cable
correctly to current lead and
power supply.
Bruker Service
Switch „Main Coil/
OFF/Shim Coil“ in
wrong position.
Change the switch position.
Bruker Service
Shims do not affect
the NMR signal.
Shim heater current
is not correct. The
shim switches are
not opened.
Set the shim heater current to
the specified value (see
Appendix Technical Data
).
Bruker Service
Содержание Ascend 600'89
Страница 1: ...Version Innovation with Integrity NMR Spectroscopy 600 89 Ascend 01 User Manual ...
Страница 4: ...4 ZTKS0174 Z31966 Rev 01 ...
Страница 7: ...7 ZTKS0174 Z31966 Rev 01 ...
Страница 8: ...8 ZTKS0174 Z31966 Rev 01 π ...
Страница 10: ...10 ZTKS0174 Z31966 Rev 01 Contact ...
Страница 34: ...34 ZTKS0174 Z31966 Rev 01 Transportation ...
Страница 35: ...35 ZTKS0174 Z31966 Rev 01 4 Assembling 4 1 Safety Approved Persons Bruker Service only ...
Страница 36: ...36 ZTKS0174 Z31966 Rev 01 Assembling ...
Страница 57: ...57 ZTKS0174 Z31966 Rev 01 8 Disassembling 8 1 Safety Approved Persons Bruker Service only ...
Страница 58: ...58 ZTKS0174 Z31966 Rev 01 Disassembling ...
Страница 59: ...59 ZTKS0174 Z31966 Rev 01 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 60: ...60 ZTKS0174 Z31966 Rev 01 Appendix ...
Страница 62: ...62 ZTKS0174 Z31966 Rev 01 Appendix ...
Страница 64: ...64 ZTKS0174 Z31966 Rev 01 Appendix ...
Страница 66: ...66 ZTKS0174 Z31966 Rev 01 Appendix ...
Страница 68: ...68 ZTKS0174 Z31966 Rev 01 Appendix ...
Страница 74: ...74 ZTKS0174 Z31966 Rev 01 Technical Data MS 600 89 Ascend Nitrogen Level Graph Figure A 3 Nitrogen Level Graph ...
Страница 75: ...75 ZTKS0174 Z31966 Rev 01 Technical Data MS 600 89 Ascend Helium Level Graph Figure A 4 Helium Level Graph ...
Страница 82: ...82 ZTKS0174 Z31966 Rev 01 Technical Data MS 600 89 Ascend This page is intentionally left blank ...
Страница 83: ...83 ZTKS0174 Z31966 Rev 01 Technical Data MS 600 89 Ascend Current Lead Figure A 10 Current Lead 55 pins ...
Страница 88: ...88 ZTKS0174 Z31966 Rev 01 Technical Data MS 600 89 Ascend This page is intentionally left blank ...
Страница 94: ...94 ZTKS0174 Z31966 Rev 01 Technical Data MS 600 89 Ascend This page is intentionally left blank ...