45
ZTKS0174 / Z31966 / Rev.: 01
Troubleshooting
Continued from page before
Indicator
Possible reason
Solution
By
The zero reading of
the Helium Level
Sensor can not be
adjusted at the
beginning of cooling
down with liquid
helium.
The Helium Level
Sensor is not
connected correctly
with the connector in
the helium flow
system.
Check the connection in the
helium fill-in turret between
Helium Level Sensor and
connector.
Bruker Service
The Helium Level
Sensor is defective.
Check the Helium Level
Sensor with the
0% calibration plug.
Bruker Service
The helium level
does not reach
100% after cooling
down.
Empty transpor-
tation dewar, helium
transfer stopped.
Refill or replace transportation
dewar.
Bruker Service
The Helium Level
Sensor is disturbed
by the transfer line’s
extension piece.
1.
Stop the liquid helium
transfer.
2.
Remove the transfer line.
3.
Measure the helium level
after some minutes
without the transfer line.
Bruker Service
After cooling down
the helium boil off is
higher than specified
(up to 5 times).
Usual behavior.
A few days are
necessary for the
radiation shields and
the insulation to
reach scheduled
temperatures.
Wait a few days and check
helium boil off.
The presence of the current
lead in the current lead turret
during energizing and
shimming helps to cool down
the radiation shield due to
higher helium flow.
Bruker Service
Содержание Ascend 600'89
Страница 1: ...Version Innovation with Integrity NMR Spectroscopy 600 89 Ascend 01 User Manual ...
Страница 4: ...4 ZTKS0174 Z31966 Rev 01 ...
Страница 7: ...7 ZTKS0174 Z31966 Rev 01 ...
Страница 8: ...8 ZTKS0174 Z31966 Rev 01 π ...
Страница 10: ...10 ZTKS0174 Z31966 Rev 01 Contact ...
Страница 34: ...34 ZTKS0174 Z31966 Rev 01 Transportation ...
Страница 35: ...35 ZTKS0174 Z31966 Rev 01 4 Assembling 4 1 Safety Approved Persons Bruker Service only ...
Страница 36: ...36 ZTKS0174 Z31966 Rev 01 Assembling ...
Страница 57: ...57 ZTKS0174 Z31966 Rev 01 8 Disassembling 8 1 Safety Approved Persons Bruker Service only ...
Страница 58: ...58 ZTKS0174 Z31966 Rev 01 Disassembling ...
Страница 59: ...59 ZTKS0174 Z31966 Rev 01 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 60: ...60 ZTKS0174 Z31966 Rev 01 Appendix ...
Страница 62: ...62 ZTKS0174 Z31966 Rev 01 Appendix ...
Страница 64: ...64 ZTKS0174 Z31966 Rev 01 Appendix ...
Страница 66: ...66 ZTKS0174 Z31966 Rev 01 Appendix ...
Страница 68: ...68 ZTKS0174 Z31966 Rev 01 Appendix ...
Страница 74: ...74 ZTKS0174 Z31966 Rev 01 Technical Data MS 600 89 Ascend Nitrogen Level Graph Figure A 3 Nitrogen Level Graph ...
Страница 75: ...75 ZTKS0174 Z31966 Rev 01 Technical Data MS 600 89 Ascend Helium Level Graph Figure A 4 Helium Level Graph ...
Страница 82: ...82 ZTKS0174 Z31966 Rev 01 Technical Data MS 600 89 Ascend This page is intentionally left blank ...
Страница 83: ...83 ZTKS0174 Z31966 Rev 01 Technical Data MS 600 89 Ascend Current Lead Figure A 10 Current Lead 55 pins ...
Страница 88: ...88 ZTKS0174 Z31966 Rev 01 Technical Data MS 600 89 Ascend This page is intentionally left blank ...
Страница 94: ...94 ZTKS0174 Z31966 Rev 01 Technical Data MS 600 89 Ascend This page is intentionally left blank ...