System Set-Up
Global User Manual
2-16
Removal of the Probe Head
1. Turn off electrical and pneumatic power to the CMM.
2. If installed, remove the probe from the probe head.
3. Support the probe head and remove the four (4) screws
securing the probe head to the Z ram (quill).
4. Separate the probe head from the Z ram (quill).
5. CAREFULLY disconnect the electrical cables between the
probe head and the Z ram (quill) and remove the probe head.
Caution: Electrical and pneumatic power to the CMM
must be turned off prior to removing the probe head.
Probe Head
Electrical Cables
Probe Head Connectors