English
I
3
Alarm 2 (Critical)
Program
Temperature
Program Alarm 2 to
desired critical set point,
which will trigger when
the chamber warms
up and reaches this
temperature.
Adjust tem-
perature by
1°C increments
using DOWN/UP
buttons.
Default: -140°C
DOWN (C)
UP (D)
Triple beep
sequence
every 10
seconds.
Pressing the
UP button
can silence
alarm.
Add LN2 immedi-
ately, or transfer
samples to avoid
compromise if no
LN2 available. Alarm
will automatically
reset when unit
temperature is at
least 5°C colder than
critical set point.
Alarm 2 (Critical)
Set On/Off
The alarm can be
enabled or disabled.
Toggle between
Off and On
using DOWN/UP
buttons.
Default: On
DOWN (C)
UP (D)
Alarms will
sound for
20 minutes
before stop-
ping.
Note LCD display
will have red back-
light, and blinking
“Alarm” indicator if
triggered.
System Alarms
Set On/Off
All alarms can be
simultaneously enabled
or disabled.
Toggle between
Off and On
using DOWN/UP
buttons.
Default: On (Bell
on LCD)
DOWN (C)
UP (D)
N/A
Note LCD display
will have red back-
light, and blinking
“Alarm” indicator if
triggered.
Standby
Program
Temperature
Power saving mode.
LCD will remain lit and
temperature visible,
while the chamber
continues to warm up
until it reaches -20°C, at
which point the unit will
power off.
Adjust
temperature by
1°C increments
using Down/Up
buttons.
Default: -120°C
DOWN (C)
UP (D)
N/A
ID
Assign system ID
3 digit alphanumeric
system ID. Assign
each CryoPod a unique
system ID code to track
units and downloadable
temperature log data.
Adjust each digit
by using DOWN/
UP buttons. Tog-
gle between each
digit by using the
SET button.
SET (B)
DOWN (C)
UP (D)
N/A
Low Mem
Set On/Off
CryoPod
™
Carrier can
store up to 100 hours of
log data. Low memory
indicates there is less
than 6 hours of record-
able space remaining.
Existing log data must
be downloaded to free
up space.
Toggle between
Off and On
using DOWN/UP
buttons.
Default: On
DOWN (C)
UP (D)
N/A
Temperature
log data can be
downloaded onto a
PC via the included
USB cable and
external CryoPod
™
Data Log software.
Time
The current time can
be programmed. This
setting is critical if you
will be utilizing the
temperature log data.
Adjust each digit
by using DOWN/
UP buttons. Tog-
gle between each
digit by using the
SET button.
Format: HH:MM
AM/PM
SET (B)
DOWN (C)
UP (D)
N/A
N/A
Date
The current date can
be programmed. This
setting is critical if you
will be utilizing the
temperature log data.
Adjust each
digit by using
DOWN/UP
buttons. Toggle
between each
digit by using
the SET button.
Format:
MM:DD:YY
SET (B)
DOWN (C)
UP (D)
N/A
N/A
Once all parameters are programmed, hold SET button (B) down for 3 seconds to save
parameter settings.