background image

Installation and Operation Manual

X-VA-1305-1306-eng-eu 
PN 541-C-029-AAG 
April, 2008

Models GT 1305 & 1306

5

Section 1:   Introduction

1-1 Description  

 

   

 

 

 

 

 

 

 

 

 The GT 1305 and GT 1306 flowmeters are low cost instruments intended 
for general metering installations where the operating pressure is within the 
limitations imposed by the use of borosilicate glass metering tubes. The 
instruments can be supplied with brass or 316SS female threaded end fittings. 

1-2 Specifications 

   

 

   

 

 

 

 

 

 

 

 

Performance 

 

Accuracy 

 

classification  

GT 1305 

 

   

 

 

•  

Standard Class 6 acc VDE/VDI 3513 

 

   

 

 

• 

Optional  Class 4 acc VDE/VDI 3513 

 

   

 

GT 1306 

 

   

 

 

• 

Standard Class 2.5 acc VDE/VDI 3513 

 

   

 

 

• 

Optional Class 1.6 acc VDE/VDI 3513 

 

Repeatability 

±0.5 % 

 

Scale length 

GT 1305

  

75 mm 

 

   

 

GT 1306 

127 mm 

 

Materials of construction 

 

End fittings  

Brass or 316 SS, fitted in aluminium collar 

 

Metering 

 

tubes   

Borosilicate glass 

 

Floats  

#7-1306

  

Glass, 316 SS or Monel 

 

   

 

#8-1305

  

PPL, Ertacital, Glass, or 316 SS 

 

   

 

#10-1305

  

316 SS 

 

   

 

#8/10-1306

  316 SS 

 

O-ring seals 

Viton, Buna or Kalrez 

 

Side plates  

Aluminum 

 

Scale type 

 

•  

Standard    Millimeter or percent of  maximum flow 

etched on metering  tube 

 

   

 

 

•  

Optional    Direct reading engraved  on a detachable 

aluminum plate 

 

Ratings 

 

 

  WARNING:Do not operate this instrument in excess of the specifications 

listed above. Failure to heed warning may result in serious personal 
injury or damage to equipment.

Section 1 Introduction 

  Meter  

Max. pressure at 

Max. size  

Pressure 

  size  

90°C(bar)  

temperature   reduction 

 

 

 

°C 

above 90 °C

 

 

GT1305         GT1306 

 

bar/°C

 

7  

-  

24  

120  

0.093 

 

8  

13.8  

20.7  

120  

0.093 

 

10  

8.9  

6.9  

120  

0.056  

Содержание GT 1305

Страница 1: ...Installation and Operation Manual X VA 1305 1306 eng eu PN 541 C 029 AAG April 2008 Models GT 1305 1306 1 Brooks O ring Seal Flowmeters Model 1306...

Страница 2: ...plied with the product Inform and educate your personnel in the proper installation operation and maintenance of the product Install your equipment as specified in the installation instructions of the...

Страница 3: ...highest quality available offering superior performance reliability and value to the user It is designed with the ever changing process conditions accuracy requirements and hostile process environment...

Страница 4: ...2 Installation 2 1 General 8 2 2 Return of equipment 8 2 3 Recommended storage practice 8 2 4 Return shipment 9 2 5 Transit precautions 9 2 6 Installation 10 Section 3 Operation 3 1 Start up 11 3 2 R...

Страница 5: ...3 Repeatability 0 5 Scale length GT 1305 75 mm GT 1306 127 mm Materials of construction End fittings Brass or 316 SS fitted in aluminium collar Metering tubes Borosilicate glass Floats 7 1306 Glass 31...

Страница 6: ...133 FE 1400 3580 8 36 8 40 0 108 0 8 66 Table 2 Flow Ranges For GT 1306 SIZE TUBE TYPE FLOAT MODEL WATER V I C 2 AIR CODE l h P kPa cSt m3n h P kPa 7 R 7M 127 1F Glass A1 31 0 25 1 1 20 0 25 316 SS A3...

Страница 7: ...with built in flow control needle valve The internal diaphragm actuated control valve is positioned by the incoming fluid pressure on one side of the diaphragm and outlet pressure spring action on the...

Страница 8: ...3569 Tel 215 362 3700 Fax 215 362 3745 E mail BrooksAm BrooksInstrument com www BrooksInstrument com Brooks Instrument Brooks Instrument Neonstraat 3 1 4 4 Kitasuna Koto Ku 6718 WX Ede Netherlands To...

Страница 9: ...Eu BrooksInstrument com Instrument must have been purged in accordance with the following All flow instruments returned to Brooks requires completion of Form RPR003 1 Brooks Instrument Decontamination...

Страница 10: ...t for damage during shipping b The indicator must be installed with the inlet at the bottom and the outlet at the top and must be installed as near vertical as possible to maintain indicator accuracy...

Страница 11: ...tion When initiating flow slowly open the valve to avoid a flow surge By pass is a help in bringing the flow on smoothly Avoid starting a pump to supply the flowmeter without the use of a valve upstre...

Страница 12: ...the polycarbonate sleeve surrounding the flow tube e Remove the float from the tube f Using a suitable solvent carefully swab and flush the inside of the metering tube Clean the float and blow dry al...

Страница 13: ...ation Should the equipment require repair contact the nearest sales service office It is important that servicing is performed only by trained and qualified service personnel If the equipment is not p...

Страница 14: ...R SIZE A B C D E F G H J SIZE 7 127 300 1 2 60 250 264 261 55 55 SIZE 8 127 300 1 2 60 250 264 261 55 55 SIZE 10 127 300 1 66 250 264 261 58 57 PANEL CUT OUT FOR BACK PANEL MOUNTING PANEL CUT OUT FOR...

Страница 15: ...Installation and Operation Manual X VA 1305 1306 eng eu PN 541 C 029 AAG April 2008 Models GT 1305 1306 15 Section 6 Exploded view and parts lists Section 6 Exploded view and part lists...

Страница 16: ...B 049 BNA 8 RV 31 316 SS FLOAT 2 1 346 B 050 BNA 8 RS 31 316 SS FLOAT 2 1 346 B 148 BNA 8 RV 8 316 SS FLOAT 2 1 346 B 149 BNA 8 RV 14 316 SS FLOAT 2 1 346 B 154 BNA 8 RS 8 316 SS FLOAT 2 1 346 B 155 B...

Страница 17: ...Kalrez CODE CONNECTION TYPE 1 NPT 2 BSP CODE SCALE TYPE A mm Decal cal curve Class 6 GT 1305 only B mm Decal cal curve Class 4 GT 1305 only C mm Decal cal curve Class 2 5 GT 1306 only D mm Decal cal c...

Страница 18: ...18 This page was intentionally left blank...

Страница 19: ...19 This page was intentionally left blank...

Страница 20: ...onger This limited warranty is the only warranty made by Seller and can be amended only in a writing signed by an authorized representative of Seller BROOKS SERVICE AND SUPPORT Brooks is committed to...

Отзывы: