Bronkhorst®
Instruction Manual ES-FLOW™ 1xxC
9.17.153A
33
Firmware Version
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[6]
R
-
105
113/5
0xF128…0xF12A/ 61737…61739
Revision number of the firmware
Identification Number
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
B
0…255
175
113/12
0x0E2C/3629
Bronkhorst® (digital) device type identification number.
Device Type
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[6]
R
-
90
113/1
0xF108…0xF10A/ 61705…61707
Device type information string; this parameter contains an abbreviation referring to the identification number.
6.1.8
Special parameters
Init Reset
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
82/64
7
0/10
0x000A/11
Init Reset
is used to unlock secured parameters (marked with a
B
symbol) for writing. It supports the following values:
Value
Description
64
unlocked, secured parameters can be read and written to
82
locked, secured parameters are read-only
At power-up,
Init Reset
is always set to ‘Locked’ (value 82).
Reset
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
R
0…7
114
115/8
0x0E68/3689
This parameter is used to reset the program, counter or alarms.
Value
Description
0
No reset
1
Reset counter
2
Reset alarm
3
Reset counter
4
Reset and disable counter
5
Reset firmware program (soft reset)
6
Reset
Alarm info
error bit
7
Reset
Alarm info
warning bit
The
Reset
parameter may be disabled by
Reset Alarm Enable
or
Reset Counter Enable
. Make sure the value is accepted
by sending value 0 first.