Bronkhorst®
Instruction Manual EL-PRESS™
9.17.101I
35
Serial Number
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[20]
R
-
92
113/3
0xF118…0xF11F/ 61721…61728
Instrument serial number for identification.
BHT Model Number
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[35]
RW
-
91
113/2
0xF110…0xF117/ 61713…61720
This parameter shows the Bronkhorst® instrument model type information.
Firmware Version
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[6]
R
-
105
113/5
0xF128…0xF12A/ 61737…61739
Revision number of the firmware
Identification Number
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char
RW
0…255
175
113/12
0x0E2C/3629
Bronkhorst® (digital) device type identification number.
Device Type
Type
Access
Range
FlowDDE
FLOW-BUS
Modbus
Unsigned char[6]
R
-
90
113/1
0xF108…0xF10A/ 61705…61707
Device type information string; this parameter contains an abbreviation referring to the identification number.
4.4
Alarms
Alarm settings are most easily accessible using FlowPlot or FlowView or a Bronkhorst® readout and control unit.
The built-in alarm functionality can be used to handle different alarm types:
·
system errors and warnings
·
min/max alarms
·
response alarms
·
batch alarms
·
master/slave alarms
The used alarm type can be set with parameter
Alarm Mode
. When an alarm is activated, the type can be read out using
parameter
Alarm Info
. An automatic setpoint change can be set using the parameters
Alarm Setpoint Mode
and
Alarm New
Setpoint
. It is also possible to set an alarm delay, to prevent overreaction to small disturbances, using parameter
Alarm Delay
Time
. The methods by which an alarm can be reset are controlled by
Reset Alarm Enable
.