MASS-STREAM
TM
Manual Flow Switch D-6300/FS
9.17.114A – edition: 2017-07
Page 23 of 36
2.6 Specifications of the MASS-STREAM
TM
Flow Switches D-6300/FS
Measurement System
Accuracy
(based on calibration with Air)
± 2% FS including non-linearity
(± 1.5% FS on request)
Repeatability
< ± 0.2% FS
Pressure sensitivity
± 0.3% RD / bar typical (Air)
Temperature sensitivity
± 0.2% RD / °C (Air)
Response time sensor (
τ
63%)
D-
631●
& D-
632●:
approx. 0.3 s
other models:
approx. 0.9 s
Leak integrity (outboard)
< 2·10
-8
mbar l/sec He
Mechanical Parts
Sensor
Stainless steel SS316 (AISI 316L)
Instrument body
Aluminium AL 50ST/51ST (anodised) or
Stainless steel SS316 (AISI 316L)
Sieves
Stainless steel SS316 (AISI 316L)
Distance parts
Teflon or Stainless steel
Protection
IP65
Operating Limits
Measuring range
up to 1…100% (1:100)
Temperature
0…50 °C
Pressure
0…10 bar(g) for aluminium body
0…20 bar(g) for stainless steel body
Warm-up time
30 min for optimum accuracy
within 30 sec for accuracy ± 4% FS
Electrical Properties
Supply voltage
+15 … 24 Vdc ± 10%
Current peak values
approx. 105 mA at 0% flow
approx. 155 mA at 100% flow
Output signal
0…10 Vdc / 0…5 Vdc active
0…20 mA / 4…20 mA active
Instrument connector
8-pin round DIN(male) for analogue and RS232