MASS-STREAM
TM
Manual Flow Switch D-6300/FS
9.17.114A – edition: 2017-07
Page 19 of 36
Calculation of the required maximum switch-point in [%]:
Switch
−
point[%] =
38.5 ln/min
45 ln/min
∙
100% = 85.6%
The
ʹM
ax
alarmʹ
must therefore be set to 85.6%, so that the instrument emits an alarm
and thus a voltage to Pin 5 when the flow rate exceeds „38.5 l
n
/min“.
•
Input of the Switch-Points via the Multi-functional display
Select
Min alarm
with
Down
and activate this menu item with
Enter
:
Down
Enter