Commissioning
B125 06/16 Rev.1
TOPAS E350V-E600V
17
7
Commissioning
7.1
Setup
The oil filling point, oil sight class and oil outlet must be easily accessible. The
cooling air inlet and the cooling air outlets must have a distance of at least 20 cm
from the adjacent walls. Emerging cooling air must not be sucked back in again.
Fig. 4 TOPAS main components
For the components, see Fig. 4 page 14.
Fig. 6 Structural diagram
For maintenance work, a minimum distance of 0.4 m must be observed in front
of the oil filter and the oil reservoir cover; see Fig. 6.
The
TOPAS
E350V
to
E600V
can only be operated error-free in a horizontal
setup location. The pumps must not be tilted. The setup of the vacuum pump on
solid ground is possible without anchorage. The vibrations from these rotary vane
vacuum pumps are minimal.
In the case of setup on a lower construction, we recommend anchorage via elas-
tic buffer elements. The vacuum pumps are equipped with four rubber shock
absorbers (lower side with threaded hole).
We recommend using these threaded holes for anchoring the pump to a base
frame.
INFORMATION!
For installations higher than 1000 m above sea level, a loss in power is
noticeable.
In this case, we ask for prior consultation with the manufacturer.
Anchorage