OTF 5000/HS/LS-001/2/3/4 Iss 4 June 2012
6
7
For /MAR - Motorised Advance Rewind (where applicable): When fitting the microtome, the
electrical socket and plug must be connected and screwed together. Before removing the
microtome from the cryostat, the electrical socket and plug must be unscrewed and
disconnected. Ensure that the power supply is disconnected before installing or removing the
microtome.
8 For /D - Cryomatic Specimen Temperature Control (where applicable): Install the
temperature heater/vee block in the microtome head first and then route the cable around
the microtome using the cable clips already fitted. Ensure that the power supply is
disconnected before installing or removing the microtome.
9 Position the left tray.
10 Finally, place the plastic brush and debris tray in position at the front of the microtome,
ensuring it is in contact with the base of the microtome.
To Remove the Microtome (Refer to Fig 4):
1
Remove Debris Tray in front of microtome.
2
Remove Advance/Rewind Knob (on top of microtome) by pulling upward.
3
Remove LH tray by sliding forward and up.
4
If a heated v-block is fitted then lift it & move it out of the way
5
Remove Coupling by loosening grub screws and slide onto Drive shaft (see Fig 4)
6
Use 2.5mm Allen Key (supplied)
7
If MAR then unplug socket at bottom Left of Tank.
8
With your right hand under the front of the microtome and your left hand under the rear
left side, carefully lift the microtome from the pins in the chamber.
1.2.6 INSULATED NIGHT PLUG
Unpack and store ready for use when required
1.2.7 /X - THIN SECTION CONVERSION KIT (where applicable)
Refer to section 2.2.3
1.2.8 /U - ULTRA VIOLET LIGHT DECONTAMINATION UNIT (where
applicable)
Unpack and store ready for use when required
1.2.9 /Y - GLOVE PORT (where applicable)
For access to the cryostat chamber, press the glove port unit into the lower half of the window
recess and then close the window until it seals against the glove port unit.
1.2.10 /V - FOOT SWITCH (where applicable)
Plug the foot switch lead into the socket on the lower front surface of cabinet, right hand side.
Place the foot switch on the floor in a convenient position.