Document # 2103017 RevA
2
Printed in USA © 2017
TROUBLESHOOTING
Operating Test
It is possible to verify normal operation of the series
6000 tables by performing the following series of
tests.
ALL MODELS
Lift test
1. Ensure ON/OFF switch is in ON position.
2. Press and hold the TABLE UP pedal. The
table should rise to the full up position
(about 37”) and stop. There must be no
continuing motor hum after the table stops.
3. Release the TABLE UP pedal.
4. Press the TABLE DOWN pedal. While
the table is going down pressing on each
corner of the back of the seat must cause
the table to momentarily stop.
5. The table should go to the full down position
Release the TABLE DOWN pedal.
6. Reach under the front section bottom plate
and lift up momentarily on each corner of
the plate. This should activate the lift causing
the table to rise while the plate is lifted.
7. Turn the ON/OFF switch, located on the
back of the rear cabinet, to the OFF position.
Verify that pressing any of the foot pedals
does not cause motion. Turn the ON/OFF
switch to the ON position.
MODELS 6500, 6501, 6800 and 6801 ONLY
Powered Back test
1. Press the TABLE UP pedal to raise the table
above 28” so the back may be lowered
without contacting the back cabinet.
2. Press the BACK DOWN pedal. The back
should go to the full recline position, which is
level.
3. Press the BACK UP pedal. The back should
go to the full up position.
4. Press the BACK DOWN pedal. While the
back is going down pressing on any corner of
the back of the seat must cause the back to
momentarily stop.
5. Press the blue RETURN TO CHAIR button on
the foot control. While the chair is in motion
momentarily press the red STOP button. The
chair must stop and remain stopped.
6. Press the blue RETURN TO CHAIR button on
the foot control. While the chair is in motion
momentarily press the BACK UP pedal. The
chair must stop and remain stopped.
7. Press the blue RETURN TO CHAIR button on
the foot control. While the chair is in motion
momentarily press the BACK DOWN pedal.
The chair must stop and remain stopped.
8. Press the blue RETURN TO CHAIR button
on the foot control. While the chair is in mo-
tion momentarily press the TABLE UP pedal.
The chair must stop and remain stopped.
9. Press the blue RETURN TO CHAIR button on
the foot control. While the chair is in motion
momentarily press the TABLE DOWN pedal.
The chair must stop and remain stopped.
10. Raise the table to the full up position, and
then lower the back so the table is in the high
table position. Press the blue RETURN TO
CHAIR button. The table should first raise the
back then lower to the chair low position.