Is the on/off switch at the back of the 

cabinet turned on?

Yes

                                                         

No

Turn on switch then 

return to step 1

Go to step 2.1.d

Remedy problem and 

return to step 1

Return to step 1

Secure connections and 

Return to step 1

Is green light on at back of control box?

Yes                                                   No

Replace the 2 control box fuses. Is light 

now on?

Yes                                                   No

Call Brewer Tech Support 888-Brewer-1.

Step 2.1 continue here

Step #

2.1

2.1

a

b

c

d

e

f

g

h

If table has a drawer warmer does the 

green drawer warmer switch illuminate 

when turned on?

Yes                                                   No

Proceed to next step if warming.

(Replace drawer warmer switch at 

your convenience). 

Return to step 1

Lamp in green switch may be burned 

out but warmer still operates. Reach 

hand into front drawer to feel if it is 

warming after 5 minutes with switch 

in either position. Is warmer warming 

up? 

Yes                                                   No

Check for loose wire connections at 

terminal block(s) if S/N is less than 

HL002100 or if S/N has a “C” at the 

end. Check that all plugs are fully 

inserted into control box and that 

securing clips are locked in place. Are 

all connections secure?

Yes                                                   No

Press foot control pedal(s). Does relay 

in control box click?

Yes                                                   No

Is table stuck in the up position?

No                                                   Yes

Front cabinet switch (1 of 6) is stuck 

closed. Unplug table power. Isolate 

faulty switch or wire using a continuity 

tester. Replace defective switch.

Fixed                       Problem remains

Check for power at wall receptacle, 

condition of power cord, and fully 

inserted plug at back of table, and 

fuse(s) at back of table. Is power 

available, connections tight and 

fuse(s) in good condition?

No                                                   Yes

Table does not go down. Control box 

relay clicks in down direction. When 

front bottom safety plate is lifted is 

there is relay click?

No                                                   Yes

6000 SERIES TABLE TROUBLESHOOTING CHART

Содержание access high-low

Страница 1: ...ument 2103017 RevA Printed in USA 2017 Troubleshooting Guide The Brewer Company LLC N88 W13901 Main Street Suite 100 Menomonee Falls WI 53051 P 1 888 Brewer 1 F 262 251 2332 www brewercompany com Brew...

Страница 2: ...ntacting the back cabinet 2 Press the BACK DOWN pedal The back should go to the full recline position which is level 3 Press the BACK UP pedal The back should go to the full up position 4 Press the BA...

Страница 3: ...f 6 not closing when pressed Press each one and listen for click when pressed and again when released Replace as necessary Fixed Problem remains Seat back switches 1 or more of 4 not opening when pres...

Страница 4: ...r 5 minutes with switch in either position Is warmer warming up Yes No Check for loose wire connections at terminal block s if S N is less than HL002100 or if S N has a C at the end Check that all plu...

Страница 5: ...ckrest movement powered back models Does motor hum in the up direction but there is no movement No Yes Backrest or lift stuck in the full up position No motor hum Do control box relays click in both u...

Страница 6: ...luminate Yes No Replace warmer outlet fuse s at back of table on plate where foot control and power cord plug in Does outlet operate Yes No Replace warmer switch Does warmer switch illuminate Yes No R...

Страница 7: ...ment not stopped by stop button or foot control pedals No Yes The backrest actuator return chair that protrudes from the center of the back of the seat section is defective or its wiring harness blue...

Отзывы: