Document # 2103017 RevA
4
Printed in USA © 2017
Is the on/off switch at the back of the
cabinet turned on?
Yes
No
Turn on switch then
return to step 1
Go to step 2.1.d
Remedy problem and
return to step 1
Return to step 1
Secure connections and
Return to step 1
Is green light on at back of control box?
Yes No
Replace the 2 control box fuses. Is light
now on?
Yes No
Call Brewer Tech Support 888-Brewer-1.
Step 2.1 continue here
Step #
2.1
2.1
a
b
c
d
e
f
g
h
If table has a drawer warmer does the
green drawer warmer switch illuminate
when turned on?
Yes No
Proceed to next step if warming.
(Replace drawer warmer switch at
your convenience).
Return to step 1
Lamp in green switch may be burned
out but warmer still operates. Reach
hand into front drawer to feel if it is
warming after 5 minutes with switch
in either position. Is warmer warming
up?
Yes No
Check for loose wire connections at
terminal block(s) if S/N is less than
HL002100 or if S/N has a “C” at the
end. Check that all plugs are fully
inserted into control box and that
securing clips are locked in place. Are
all connections secure?
Yes No
Press foot control pedal(s). Does relay
in control box click?
Yes No
Is table stuck in the up position?
No Yes
Front cabinet switch (1 of 6) is stuck
closed. Unplug table power. Isolate
faulty switch or wire using a continuity
tester. Replace defective switch.
Fixed Problem remains
Check for power at wall receptacle,
condition of power cord, and fully
inserted plug at back of table, and
fuse(s) at back of table. Is power
available, connections tight and
fuse(s) in good condition?
No Yes
Table does not go down. Control box
relay clicks in down direction. When
front bottom safety plate is lifted is
there is relay click?
No Yes
6000 SERIES TABLE TROUBLESHOOTING CHART