System Maintenance
52
Section 6 Maintenance
ACIST Medical Systems, Inc.
P/N 900468-001 Rev. 02
To remove the chamber sleeve:
1. From an open position, gently pull down on the sleeve until it
pops out from its guides.
2. Detach the cable from the system.
If you wish to clean the chamber sleeve by holding it under a
faucet, be sure you hold the attached cable a safe distance
from the water.
CAUTION:
Do not immerse or expose the cable to water.
Electrical connections from the syringe valve
sensor are attached to the chamber and may
produce “Syringe Valve Open/Closed” messages if
the cable is damaged by exposure to water.
Do not allow water to come in
contact with the cable during
cleaning
Cleaning the Removable Backlight Cover
ACIST injection systems manufactured after mid-2003 have a
removable backlight cover that should be removed and cleaned at
the end of each day.
To clean the removable backlight cover:
1. Turn off the power.
Make sure the power is off whenever you remove the backlight
cover or clean the system.
2. Remove the backlight cover.
Be sure that the air column detect sensor door is not latched,
so that it will not catch on the removable cover. Carefully
pull the backlight cover away from the outside edge of the
backlight, pulling all the way around the perimeter of the cover.
Lift the cover off the syringe chamber, allowing the manifold
sensor and the air column detect sensor to go through the
appropriate cutout holes.
3. Clean the backlight cover and the backlight area using a soft
cloth and warm water.
If decontamination is necessary, you can use a bactericide
on the removable backlight cover, but use only warm water
to clean the backlight area and the rest of the ACIST injection
system.
4. Reattach the backlight cover.
Position the backlight cover over the backlight, lining up the
cutouts for the manifold sensor and the air column detect
sensor. Be sure the air column detect sensor door is positioned
out of the way so it does not catch on the backlight cover.
Secure the backlight cover by snapping the outer edge of
the backlight into the “channel” around the edge of the cover.
Start by snapping the cover into the upper-right corner of the
backlight and work all around the perimeter of the backlight
cover until the entire cover is securely in place.
Turn off the power
Remove the backlight cover and
clean with a soft cloth and warm
water
Reattach the backlight cover
Cleaning the Removable Backlight Cover
Содержание ACIST CVi CMS2000
Страница 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Страница 2: ...This page intentionally left blank...
Страница 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Страница 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...