Performing Patient Procedures
42
Section 5 Performing Procedures
ACIST Medical Systems, Inc.
P/N 900468-001 Rev. 02
3. Purge the high pressure (injection) tubing and stopcock with
contrast to clear the saline from tubing. In using a pressure
transducer, ensure that it has been set up properly as
described in the “System Setup “ section.
4. Connect the 3-way stopcock to the patient catheter. Use good
clinical practices to ensure there is no air in the tubing.
5. If using a CL100H, CMS2000 or CVi in cardiac mode, enter the
patient’s weight.:
- Press PATIENT WEIGHT to display the keypad.
- Type the patient’s weight (in kg) and then press Enter.
The settings for the parameters (
fl
ow, volume, pressure, and
rise time) will be calculated automatically by the system.
6. Select injection type by pressing the desired selection on the
touchscreen. Injection types include:
Cardiology (CL100H,CMS2000,
CVi - in cardiac mode)
LCA (left coronary artery)
RCA (right coronary artery)
LV/Ao (left ventricle, aorta)
Other
Peripheral (Voyager, CVi-in
peripheral mode)
Pigtail
Selective
Microcatheter
Other
When a new injection type is selected, default parameters will
be displayed.
7. Review and revise injection parameters as clinically indicated
for the patient.
The system is now ready for patient injections.
Purge the high pressure
(injection) tubing and stopcock
Patient Weight Button
Injection Type Buttons
Pre-Procedure T
asks
Содержание ACIST CVi CMS2000
Страница 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Страница 2: ...This page intentionally left blank...
Страница 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Страница 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...