Information & W
arnings
P/N 900468-001 Rev. 02
ACIST Medical Systems, Inc.
xi
Precautions
LOCKING WHEELS
After moving the ACIST system using the pedestal cart, lock the wheels to prevent
unintentional movement when the cart is stationary.
SALINE PUMP
The tubing must be properly installed in the injector head and the locking V-teeth
engaged on the tubing for proper operation of the pump and system. (See page 28)
PRESSURE TRANSDUCER (OPTIONAL)
Attach the pressure transducer cartridge to the transducer backplate before
application of any positive pressure to the system. This prevents pressures from
bursting the dome membrane and introducing air into the system.
PREVENTATIVE MAINTENANCE
To ensure that your ACIST system is in optimal working condition, annual
preventative maintenance is recommended. Contact ACIST Medical Systems for
information on extended warranty options (see page 73).
PROPERTIES OF CONTRAST
For correct function of the ACIST system, make sure that the contrast has its
viscosity maintained between 26.6 centipoise and 4.6 centpoise for all functions at
the temperature used.
THE MOUNTED INSTRUMENT
Never lean, grab or place objects on the ACIST System. When transporting the
system, guide it using the pedestal cart handrail only. Do not grab or push on the
system itself. Make sure safety latch knob is tightened in the clockwise rotation and
the unit is secure on the cart. For power supplies that are off the patient table (bed)
mount, be sure that the power supply is in the cart tray during transportation.
TRAINING
ACIST Medical systems recommends instruction for all quali
fi
ed persons prior to
operating of the ACIST system. A certi
fi
ed ACIST Medical systems representative
will conduct training.
Содержание ACIST CVi CMS2000
Страница 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Страница 2: ...This page intentionally left blank...
Страница 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Страница 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...