Imaging System Status and Control Lines
The imaging system (or x-ray) interface is a general group of
status and control lines that allow for a wide variety of interface
con
fi
gurations with
fi
lm changers, programmers, or digital imaging
systems. The lines are described in the following tables.
Status Output Lines
Injecting
This output is active anytime the
injection system is injecting contrast.
Armed Status
This output is active while the injection
system is armed and ready for an
injection.
X-ray trigger out
This trigger signal is set
as active by the injector
as a trigger to the
imaging system.
There are several conditions that must
be met before this signal is active.
1. The injector must be armed.
2. The injector start must be active
3. The X-ray delay has passed.
Hand Controller Pressed
(only for CVi standard
and Voyager-Philips
Alura FD10 and FD 20)
There are several conditions that must
be met before this signal is active.
1. The injector must be armed.
2. The hand controller disable is
active.
3. The hand controller contrast button
has been pressed.
Input Control Lines
Disarm/Disable
When this signal is set active from an
external device, the injection system
will disarm (if armed), and will be
prevented from arming or injecting until
the signal is cleared.
Injector Start
When this input is active, starts the
programmed injection process.
Hand Control Disable
Disarms the hand controller inject
function when X-ray Sync is ON.
Imaging System Status and Control Lines
Содержание ACIST CVi CMS2000
Страница 1: ...ACIST Contrast Delivery System User Manual 900468 001 02 2010 01 and CMS2000 and E2000Voyager...
Страница 2: ...This page intentionally left blank...
Страница 14: ...Information Warnings xii ACIST Medical Systems Inc P N 900468 001 Rev 02...
Страница 91: ...P N 900468 001 Rev 02 ACIST Medical Systems Inc Notes 77 Notes...