7 Preparation
MN031-543-S0 EN
TissueSeal / TissueSeal PLUS COMFORT
29
Jaw insert
1.
Check the ratchet
B1
for ease of motion.
2.
Check the jaw holder
B2
for damage and corrosion.
Jaws
1.
Check the electrode surfaces
A2
for cleanliness.
2.
Check the gap between the electrode surfaces
A2
in the closed
position for uniformity.
3.
Check the electrode surfaces
A2
for damage.
HF cable
1.
Check the connector for damage and corrosion.
2.
Visually inspect the insulation for damage.
7.6.
Packing
The
packaging
must meet the following requirements:
EN (ANSI AAMI) ISO 11607/ EN 868-2...10 (formerly EN
868/ANSI AAMI ISO 11607)
Suitable for steam sterilization (resistant to temperatures up to
137 °C, with sufficient steam permeability)
Serviced at regular intervals (sterilization container)
Pack the ligation instrument in a suitable disposable sterilization
package and/or a suitable sterilization container.
Sterilization in the transportation packaging is not allowed.
Содержание TissueSeal COMFORT
Страница 1: ...DE Gebrauchsanweisung TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 2: ......
Страница 35: ......
Страница 37: ...EN Operating Manual TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 38: ......
Страница 39: ...Legend MN031 543 S0 EN TissueSeal TissueSeal PLUS COMFORT 3...
Страница 40: ...4 TissueSeal TissueSeal PLUS COMFORT MN031 543 S0 EN...
Страница 73: ......
Страница 75: ...FR Mode d emploi TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 76: ......
Страница 110: ......
Страница 112: ...IT Instruzioni per l uso TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 113: ......
Страница 146: ...MN031 543 S0 IT TissueSeal TissueSeal PLUS COMFORT 35...
Страница 148: ...ES Manual de instrucciones TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 181: ...10 Descripci n general del sistema 34 TissueSeal TissueSeal PLUS COMFORT MN031 543 S0 ES...
Страница 183: ...PL Instrukcja obs ugi TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 184: ......