7 Preparation
24
TissueSeal / TissueSeal PLUS COMFORT
MN031-543-S0 EN
7.2.
Dismantling
Dismantle the ligation instrument (see section 6, page 19).
7.2.1.
Pretreatment in an ultrasonic bath
CAUTION
Risk of infection due to water spray and vapors from the
ultrasonic bath!
Wear a face mask and protective clothing.
Adequate ventilation is recommended.
1. Place the ligation instrument in the ultrasonic bath for at least 5 minutes.
Position instrument parts with large surface areas in the ultrasonic bath
such that they will not be damaged by the ultrasonic energy.
2. Use suitable cleaning and disinfection products for ultrasonic cleaning
(see section 7.11, page 30).
3.
Follow the manufacturer’s instructions with regard to the concentration
and exposure time of the cleaning and disinfection products.
Содержание TissueSeal COMFORT
Страница 1: ...DE Gebrauchsanweisung TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 2: ......
Страница 35: ......
Страница 37: ...EN Operating Manual TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 38: ......
Страница 39: ...Legend MN031 543 S0 EN TissueSeal TissueSeal PLUS COMFORT 3...
Страница 40: ...4 TissueSeal TissueSeal PLUS COMFORT MN031 543 S0 EN...
Страница 73: ......
Страница 75: ...FR Mode d emploi TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 76: ......
Страница 110: ......
Страница 112: ...IT Instruzioni per l uso TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 113: ......
Страница 146: ...MN031 543 S0 IT TissueSeal TissueSeal PLUS COMFORT 35...
Страница 148: ...ES Manual de instrucciones TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 181: ...10 Descripci n general del sistema 34 TissueSeal TissueSeal PLUS COMFORT MN031 543 S0 ES...
Страница 183: ...PL Instrukcja obs ugi TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 184: ......