5 Operation
20
TissueSeal / TissueSeal PLUS COMFORT
MN031-543-S0 EN
5.
Using the foot switch of the HF device, activate the HF current for
coagulation:
A continuous acoustic signal sounds during the entire
sealing process to indicate that power is being supplied.
An alternating acoustic signal indicates the end of the
sealing process.
6.
Release the foot switch to stop the supply of power.
7.
Using the handle
B
, open the jaws
A
.
The tissue is coagulated.
Changing the output power of the HF device
Before increasing the output power of the HF device, check:
all HF cables and connectors are properly connected;
the ligation instrument is properly connected (see the
operating instructions of the HF device);
the foot switch works properly;
the insulation of the HF cable,
the active electrodes of the jaws
A
are clean and not worn
out.
5.3.
Withdrawal
WARNING
Risk of patient injury due to damaged or broken-off
parts!
Check the ligation instrument after each use. All parts
must be present.
Содержание TissueSeal COMFORT
Страница 1: ...DE Gebrauchsanweisung TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 2: ......
Страница 35: ......
Страница 37: ...EN Operating Manual TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 38: ......
Страница 39: ...Legend MN031 543 S0 EN TissueSeal TissueSeal PLUS COMFORT 3...
Страница 40: ...4 TissueSeal TissueSeal PLUS COMFORT MN031 543 S0 EN...
Страница 73: ......
Страница 75: ...FR Mode d emploi TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 76: ......
Страница 110: ......
Страница 112: ...IT Instruzioni per l uso TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 113: ......
Страница 146: ...MN031 543 S0 IT TissueSeal TissueSeal PLUS COMFORT 35...
Страница 148: ...ES Manual de instrucciones TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 181: ...10 Descripci n general del sistema 34 TissueSeal TissueSeal PLUS COMFORT MN031 543 S0 ES...
Страница 183: ...PL Instrukcja obs ugi TissueSeal PLUS COMFORT TissueSeal COMFORT...
Страница 184: ......