8 Maintenance and Repair
900-001_IFU_V1.0_11270-S0-20130118-EN
ARC PLUS Operating Manual
37
8.
Maintenance and Repair
8.1.
Maintenance
DANGER
Infection hazard
To avoid spreading germs and infections, disinfect the device and pack it in
the shipping packaging before it leaves the hospital or surgical practice.
After each use, check the device, the device trolley and the accessories
(e.g. foot switch and cable) for damage or defects. In particular, ensure that
the insulation is intact on all cables.
Do not use any damaged device, damaged device trolley or damaged
accessories.
Replace defective accessories immediately.
Have the safety inspection for the device performed once a year. Consult the
appropriate service manual for additional technical information.
8.1.1. Technical safety inspection (TSI)
Safety inspections must be performed once a year.
The device and accessories may be inspected only by persons who have the
required training, knowledge and experience and who can perform the
inspection independently.
With regard to the safety inspection, you must comply with the country-
specific rules and regulations.
The inspector documents the inspection results and measured values according to
the inspection record.
In the case of severe deviations from the values on the final acceptance report, or
if the stated maximum values have been exceeded:
Send the ARC PLUS to a service centre (see Section “Technical service“,
page 39).