6 Detecting and Correcting Faults
900-001_IFU_V1.0_11270-S0-20130118-EN
ARC PLUS Operating Manual
35
Heading
Fault message
ARC PLUS
Warning
Please check if the instrument is free of adhesions, and purge it with
argon.
If repeated purging does not solve the problem, the instrument and
cable must be replaced.
ARC PLUS
Fault
Please check if the argon bottles are connected and open. Empty bottles
should be replaced.
You can connect two argon bottles. The unit shifts automatically to the
second bottle
ARC PLUS
Warning
The argon flow settings at ARC 400 are invalid
ARC PLUS
Warning
The filling level of the argon bottle is low.
Please make sure that a replacement is available.
You can connect two argon bottles. The unit shifts automatically to the
second bottle.
ARC PLUS
Fault
The argon bottle is empty.
Connect a replacement bottle to enable activation.
You can connect two argon bottles. The unit shifts automatically to the
second bottle
ARC PLUS
Notice
The argon bottle is empty. The unit has shifted automatically to the
replacement bottle.
Please make sure that a replacement is available
ARC PLUS
Fault
Please check if the argon bottles are connected and open. Empty bottles
should be replaced. Subsequently restart ARC PLUS by activating the
flashing "Purge" button
TSI ARC
PLUS
Warning
The annual Technical Safety Inspection (TSI) for ARC PLUS is due.
Initiate this inspection as soon as possible.
Any further use of the unit is at the user's risk.
Contact the service centre (see Section “Repairs”. page 38) if you are unable
to correct a fault by following the indicated instructions.
Unlisted faults
Additional detailed information is available in the Service Manual.
Contact the service centre (see Section “Repairs”. page 38) in case of any
faults not shown in the fault list.
If the expected change in the tissue does not occur and no fault message
appears, check the parameter values and the accessory connections.