12 EMC
900-001_IFU_V1.0_11270-S0-20130118-EN
ARC PLUS Operating Manual
43
12.
EMC
12.1. Guidelines and manufacturer's declaration in
accordance with IEC 60601-1-2, Section 6.8.3.201
Electromagnetic interference (IEC 60601-1-2, Table 201)
The ARC PLUS is intended for operation in an electromagnetic environment as described below. The
customer or user of the ARC PLUS should ensure that it is operated in such an environment.
Interference measurements
Conformity
Electromagnetic environment
guideline
HF emissions as specified in CISPR
11
Group 2
The ARC PLUS must emit
electromagnetic energy in order to
perform its intended functions.
Nearby electronic devices may be
affected.
HF emissions as specified in CISPR
11
Class B
The ARC PLUS is suitable for use
in facilities other than those in the
residential environment and those
connected directly to a public
power supply network that also
supplies power to buildings used
for residential purposes.
Emission of harmonics as specified
in IEC 61000-3-2
Class A
Emission of voltage fluctuations or
flicker as specified in IEC 61000-3-3
Conforms
Electromagnetic immunity (IEC 60601-1-2, Table 202)
The ARC PLUS is intended for operation in the electromagnetic environment described below. The
customer or user of the ARC PLUS should ensure that it is operated in such an environment.
Immunity test
IEC 60601 test level
Conformity level
Electromagnetic
environment guidelines
Electrostatic discharge
(ESD) according to
IEC 61000-4-2
±6 kV contact
discharge
±6 kV contact
discharge
Floors should be made of
wood or cement or
covered with ceramic
tiles. If the floor is
covered with synthetic
material, the relative
humidity must be at least
30%.
±8 kV air discharge
±8 kV air discharge
Fast transient electrical
interference or bursts as
specified in
IEC 61000-4-4
±2 kV on power supply
lines
±2 kV on power supply
lines
The quality of the mains
power should correspond
to that of a typical
business or hospital
environment.
±1 kV on input and
output lines
±1 kV on input and
output lines
Surges as specified in
IEC 61000-4-5
±1 kV between
external lines
±1 kV between external
lines
The quality of the mains
power should correspond
to that of a typical
business or hospital
environment.
±2 kV external line to
ground
±2 kV external line to
ground