![BOREAL LASER GasFinderAB Скачать руководство пользователя страница 46](http://html.mh-extra.com/html/boreal-laser/gasfinderab/gasfinderab_operation-manual_2788301046.webp)
9.6
Boreal Laser Inc.
GasFinderAB Operation Manual — Appendices
July 17th 2017
Explanation of R
2
(continued)
A typical plot of
concentration
versus
R
2
will give the following graph:
With lower levels of sample gas, the
R
2
s
decrease, and equal zero when there is
no gas present. As the signal from the gas
becomes stronger, the effect of noise, both
electronic and optical, is reduced and the
R
2
s will increase (i.e., the signal to noise
ratio will increase). The general shape of the
plot is the same for all gases; however, the
x-axis values will depend on the sensitivity
of the instrument to the gas species being
observed.