7 | QUICK MANUAL

QUICK MANUAL

ENDO P® SYSTEM

© 2021 Bonvisi® AB 
DOC ID: ART 19-005 EN ver 3

Manufactured by:

BONVISI AB
c/o IOffice
Kungsgatan 60
111 22 Stockholm, Sweden

+46 (0)8 661 50 20
www.bonvisi.com
[email protected]

2797

This product is distributed locally by

(Space for information about the authorized Bonvisi® representative) 

USING THE FOOTPEDAL DURING SURGERY 

MODE

SINGLE PRESS

DOUBLE PRESS

3-SECOND PRESS

Blue 

pedal

Standby

Starts Surgery

Surgery

1.  First single press     

decreases flow by 30%.

2.  Second single press 

decrease flow by further 
15%.

1.  First double press     

increases flow by 30% 
for 2 minutes.

2.  Second double press 

increases flow by further 
15% for 2 minutes.

Goes to Standby

Red 

pedal

Standby

Starts Surgery

Surgery

1.  First single press    

decreases pressure by 
30%.

2.  Second single press 

decrease pressure              
by further 15%.

1.  First double press        

increases pressure by 
30% for 2 minutes.

2.  Second double press     

increase pressure 
by further 15% for 2 
minutes.

Note: Default pressure is 
set to 30 mmHg for most 
surgery types.

Goes to Standby

Both pedals

simultaneously

Surgery

Flushing function

at 600 mL/min

Содержание ENDO P SYSTEM

Страница 1: ...cm below the patient s median line and approximately 50 cm from the ground When placed on the Bonvisi Tower shown below the ENDO P will be 50 cm from the ground which is ideal Place the ENDO H on the...

Страница 2: ...ft Check that the metallic key Fig 1c on the cassette holder is in the horizontal position then slide the cassette into place over the key until it clicks into place and is in contact with the front o...

Страница 3: ...o the rear of both devices 5 Plug the ENDO H power cord into the earthed outlet 6 Turn ON power by using the switch located on the rear of the unit In a few seconds the power on LED green and heating...

Страница 4: ...NDO P pump Personnel not wearing sterile scrubs 1 If a used check valve remains from a previous procedure this must be removed from the Day Cassette tubing see Fig 3a 2 Connect the Irrigation Tube to...

Страница 5: ...n Tube at the Luer lock beneath the check valve Fig 4b Leaving the check valve in place protects the system from contamination when the system is inactive between cases Discard the rest of the Irrigat...

Страница 6: ...inuously shown on the device display If not using the ENDO BP the device will use the procedure specific default settings Each surgery type has default settings pre set for flow and pressure If the us...

Страница 7: ...1 First single press decreases flow by 30 2 Second single press decrease flow by further 15 1 First double press increases flow by 30 for 2 minutes 2 Second double press increases flow by further 15...

Отзывы: