5 | QUICK MANUAL
QUICK MANUAL
ENDO P® SYSTEM
DISCONNECTION OF SINGLE-USE MATERIALS AFTER SURGERY
WHEN MULTIPLE SURGERIES WILL BE PERFORMED ON THE SAME DAY
Personnel not wearing sterile scrubs, use aseptic technique:
1. The connections to the Heater Bag should be left in place. No adjustments needed.
2. Close all tube clamps on the Irrigation Tube, Patient Cassette and Day Cassette. Ensure that the white
clamp on the Day Cassette´s tube furthest to the right Fig. 4a is closed.
3. Disconnect the Irrigation Tube and Patent Cassette from the operating instrument.
4. Remove the Patient Cassette from the ENDO P® by opening the cover and releasing the metallic key.
5. Disconnect the Irrigation Tube at the Luer lock beneath the check valve Fig. 4b. Leaving the check valve
in place protects the system from contamination when the system is inactive between cases. Discard the
rest of the Irrigation Tube in accordance with hospital waste procedures.
6. A new Irrigation Tube can be connected when the next surgery starts. Follow Patient Cassette instructions.
WHEN THE FINAL SURGERY OF THE DAY HAS BEEN COMPLETED
Personnel not wearing sterile scrubs:
1. Turn off the ENDO P®. Turn off the ENDO H® and wait 2 min for the hot surfaces to cool down.
2. Close all disposables’ tube clamps and disconnect all tubing from the operating instrument.
3. Remove the Patient Cassette and Day Cassette by releasing the respective metallic key 90° to the left
and opening the covers.
4. Remove the ENDO H® Heater bag by opening the ENDO H®.
5. Discard all disposables according to hospital waste policy.
4b
4a
Figure 4