5 | QUICK MANUAL

QUICK MANUAL

ENDO P® SYSTEM

DISCONNECTION OF SINGLE-USE MATERIALS AFTER SURGERY 

WHEN MULTIPLE SURGERIES WILL BE PERFORMED ON THE SAME DAY

Personnel not wearing sterile scrubs, use aseptic technique:

1.  The connections to the Heater Bag should be left in place. No adjustments needed.
2.  Close all tube clamps on the Irrigation Tube, Patient Cassette and Day Cassette. Ensure that the white 

clamp on the Day Cassette´s tube furthest to the right Fig. 4a is closed.

3.  Disconnect the Irrigation Tube and Patent Cassette from the operating instrument.
4.  Remove the Patient Cassette from the ENDO P® by opening the cover and releasing the metallic key. 
5.  Disconnect the Irrigation Tube at the Luer lock beneath the check valve Fig. 4b. Leaving the check valve 

in place protects the system from contamination when the system is inactive between cases. Discard the 

rest of the Irrigation Tube in accordance with hospital waste procedures.

6.  A new Irrigation Tube can be connected when the next surgery starts. Follow Patient Cassette instructions.

WHEN THE FINAL SURGERY OF THE DAY HAS BEEN COMPLETED

Personnel not wearing sterile scrubs:

1.  Turn off the ENDO P®. Turn off the ENDO H® and wait 2 min for the hot surfaces to cool down.
2.  Close all disposables’ tube clamps and disconnect all tubing from the operating instrument.
3.  Remove the Patient Cassette and Day Cassette by releasing the respective metallic key 90° to the left 

and opening the covers.

4.  Remove the ENDO H® Heater bag by opening the ENDO H®.
5.  Discard all disposables according to hospital waste policy.

4b

4a

Figure 4

Содержание ENDO P SYSTEM

Страница 1: ...cm below the patient s median line and approximately 50 cm from the ground When placed on the Bonvisi Tower shown below the ENDO P will be 50 cm from the ground which is ideal Place the ENDO H on the...

Страница 2: ...ft Check that the metallic key Fig 1c on the cassette holder is in the horizontal position then slide the cassette into place over the key until it clicks into place and is in contact with the front o...

Страница 3: ...o the rear of both devices 5 Plug the ENDO H power cord into the earthed outlet 6 Turn ON power by using the switch located on the rear of the unit In a few seconds the power on LED green and heating...

Страница 4: ...NDO P pump Personnel not wearing sterile scrubs 1 If a used check valve remains from a previous procedure this must be removed from the Day Cassette tubing see Fig 3a 2 Connect the Irrigation Tube to...

Страница 5: ...n Tube at the Luer lock beneath the check valve Fig 4b Leaving the check valve in place protects the system from contamination when the system is inactive between cases Discard the rest of the Irrigat...

Страница 6: ...inuously shown on the device display If not using the ENDO BP the device will use the procedure specific default settings Each surgery type has default settings pre set for flow and pressure If the us...

Страница 7: ...1 First single press decreases flow by 30 2 Second single press decrease flow by further 15 1 First double press increases flow by 30 for 2 minutes 2 Second double press increases flow by further 15...

Отзывы: