6 | QUICK MANUAL
QUICK MANUAL
ENDO P® SYSTEM
PRIMING THE IRRIGATION TUBE AND
OPERATING INSTRUMENT
• Make sure all clamps are open (one
irrigation bag at a time).
• Make sure that the instruments inflow
and outflow ports are open.
• Hold the instrument over the waste
container and press
Prime
.
• Once fluid reaches the tip of the
instrument the procedure can begin.
• You can also use
Start Surgery
(see next chapter) when priming the Irrigation Tube and instrument.
SELECTION OF PROCEDURES
When using the ENDO BP™ and its data, the control of the target pressure is dynamic and automatically
maintained based on the patient's systemic blood pressure. The patient's measured systemic blood
pressure (systolic/diastolic) and delivery pressure (DP) are continuously shown on the device display.
If not using the ENDO BP™ the device will use the procedure specific default settings. Each surgery
type has default settings (pre-set) for flow and pressure. If the use of the ENDO BP™ is interrupted or
its readings are out of range the device will automatically use the procedure-specific default settings.
Check that all clamps are open. Select the desired surgery type/procedure. Confirm the procedure
by tapping
Enter
then reconfirm by pressing
Accept
. Press
Start Surgery
when you are ready to
start the procedure.
Figure 8
SURGERY TYPE
ENDO BP™ MAY BE USED
DEFAULT FLUID FLOW
DEFAULT FLUID PRESSURE
TUR-P: PROSTATE
SURGERY
Yes
200 mL/min
50 mmHg
TUR-B: BLADDER
SURGERY
Yes
150 mL/min
30 mmHg
URETER: SURGERY IN
THE URETER
No
200 mL/min
30 mmHg
KIDNEYS: SURGERY IN
THE RENAL PELVIS
No
15 mL/min
30 mmHg
CYSTOSCOPY:
DIAGNOSTIC
EXAMINATION OF THE
URINARY TRACT
No
200 mL/min
Max 50 mmHg