4 | QUICK MANUAL
QUICK MANUAL
ENDO P® SYSTEM
CONNECTING THE PATIENT CASSETTE
Personnel wearing sterile scrubs:
1. Remove the Irrigation Tube from the Patient Cassette packaging. Close the clamp on the Irrigation Tube
and check that the Luer connectors are securely tightened. Remove the transparent cap from the Luer
connector and connect it to the inflow channel on the instrument. If the Luer connector does not fit, it
can be cut off using sterile scissors and the tubing pushed into position on the inflow port. Now pass the
Irrigation Tube over to the nonsterile (circulating) staff for connection to the Day Cassette.
2. Remove the Patient Cassette from the packaging. Remove the transparent cap from the red Luer connector
and connect it to the outflow port on the instrument. If the Luer connector does not fit, it can be cut off
using sterile scissors and the tubing pushed into position on the outflow port. Now pass out the Cassette
over to the nonsterile (circulating) staff for connection to the ENDO P® pump.
Personnel not wearing sterile scrubs:
1. If a used check valve remains from a previous procedure, this must be removed from the Day Cassette
tubing, see Fig. 3a.
2. Connect the Irrigation Tube to the blue Luer connector furthest to the right on the Day Cassette. See Fig. 3b.
3. Open the light-grey cover Fig. 3c of the right pump by moving it to the right. Check that the metallic key
on the cassette holder Fig. 3d is in the horizontal position. Slide the cassette into place over the key until
it clicks into place and is in contact with the front of the Endo P®. Secure the cassette to the pump by
turning the key 90° to the right.
4. It is important to place the tubing loop centred over the black wheel on the right pump Fig. 3e. Close the
light-grey cover by moving it to the left.
5. The tubes that run from the cassette must run freely down from the cassette without being twisted.
6. Check that all Luer locks are securely tightened. Open all clamps.
7. Place the tube labelled WASTE into a non-conductive waste container. Make sure that nothing
blocks the flow and do not connect any active device such as suction pumps to the tubing of the
ENDO P® disposables.
(!)
3a
3b
3c
3d
3e
Figure 3