Page | 16
Invalid Instrument
• A
single-pulsed
alert tone
• Instrument receptacle display
illuminates and remains
red
• Generator will not allow
RF energy delivery
An unusable instrument has been
connected.
1. Disconnect instrument from
the generator.
2. Ensure instrument is
CoolSeal
™
compatible
3. Reconnect instrument to the
generator.
4. Confirm the instrument
receptacle display
illuminates green
If the Instrument Error reoccurs:
• Do not use the instrument
• Use a different CoolSeal
™
Instrument.
Possible use conditions include:
Instrument connected is not
CoolSeal
™
technology compatible.
Confirm CoolSeal
™
technology
compatibility of the instrument
from the instrument instructions
for use.
Instrument connected has been
used previously.
Discard instrument.
Instrument is not usable with
software version.
Confirm the software version
required by the instrument is not
greater than the software version
labelled on the bottom of the
CoolSeal
™
Generator.
For a software upgrade refer to
the Software Upgrade section.