Bolder Surgical
en - CoolSeal™ Generator
Page | 31
Recommended Separation Distances Between Portable and Mobile RF Communications Equipment and the
CoolSeal
™
Generator
The CoolSeal
™
Generator is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the CoolSeal
™
Generator can help prevent electromagnetic interference by
main¬taining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
CoolSeal
™
Generator as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
Separation distance from transmitter
d = [2]√P
0.01
0.20
0.1
0.63
1
2.00
10
6.32
100
20.00
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects, and people.