5.0  Technical Specifications 

 

This equipment is for indoor use only and will meet its performance 
figures  within  the  ambient  temperature  range  of  0°C  to  40°C,  with 
maximum  relative  humidity  of  80%  non-condensing.    Installation 
category II (transient voltages).  Pollution degree 2 in accordance 
with IEC 664.  For operation at altitudes of up to  2000 meters (6500 
feet). 
 

Ambient Operating 
Temperature 

0°C to 40°C 

Tilt Range 

0° to 12° 

Load Capacity 

9 Kg (20 lb) 

Supply Ratings 

Model 281111: 115 VAC ± 10%, 0.73 A, 60 Hz 
Model  281111-2:  230  VAC  ±  10%,,  0.7  A, 
50/60 Hz 

Power Rating 

Model 281111: 50 W 
Model 281111-2: 60 W 

Rotational Speed 

Variable from 5 to 120 rpm 

  

This product has been tested to the requirements of CAN/CSA - 

 

later version of the same standard incorporating the same  

 

 

level of testing requirements. 

6.0  Accessories 

 

Model Number 

Description 

260312 

Double Stacking Tray 

C2202208 

Dimpled Rubber Mat 

 

7.0  Fault Diagnosis 

 

Symptom 

Possible Cause 

Action Required 

Unit does not operate 

a.

 

Unit not switched on 

b.

 

Unit not plugged 
into power supply 

c.

 

Fuses blown 

 
d.

 

Power supply failure 

a.

 

Switch on 

b.

 

Plug in, switch on 

 

c. Replace fuses (see 

section 8.2) 

d. Check that other 

electrical 
appliances on the 
same circuit are 

Содержание Orbitron V 281111

Страница 1: ...Boekel Orbitron V Models 281111 281111 2 Operating Instructions N2400331 Rev 2 01 2012...

Страница 2: ...5 3 2 Installation 5 4 0 Operation 5 4 1 Identification of Controls See Figure 1 5 4 2 Timer Operation 6 4 3 Tray Angle Adjustment See Figure 2 6 5 0 Technical Specifications 7 6 0 Accessories 7 7 0...

Страница 3: ...3 Figure 1 Rotator Front Panel Power Switch Speed Control Timer Control...

Страница 4: ...ement Connect only to a power supply that provides a safety ground terminal Before moving disconnect at the power supply socket Do not remove the plug from the rear of the unit Ensure that the power s...

Страница 5: ...n V Power line cord One 3 16 Hex key for tilt adjustment Operating instructions 3 2 Installation Place the Orbitron V on a clean flat and stable work surface to ensure that the suction feet on the bas...

Страница 6: ...Remove load from Tray Rotate Speed Adjustment Knob counterclockwise until the Tray is rotating slowly When the pivot block is easily accessible for adjustment turn the Power Switch off Insert the supp...

Страница 7: ...0 Hz Model 281111 2 230 VAC 10 0 7 A 50 60 Hz Power Rating Model 281111 50 W Model 281111 2 60 W Rotational Speed Variable from 5 to 120 rpm This product has been tested to the requirements of CAN CSA...

Страница 8: ...ducts are designed to comply with IEC1010 1 No routine maintenance is required 8 1 Cleaning Disengage power cord prior to cleaning If a spill occurs use appropriate clean up procedures as required for...

Страница 9: ...5mm x 20mm quick acting rated 250V for 230V units 5mm x 20mm time delay rated 250V for 115V units Model 281111 115 V AC 1 25AT Model 281111 2 230 V AC 1 00AF Push the fuse drawer back in Reconnect un...

Страница 10: ...A Decontamination Certificate must be completed signed by the user and returned to Boekel Scientific prior to receiving the RMA number Please be sure to mark the outside of the returned goods package...

Отзывы: