9
Figure 3
Fuse
Drawer
Figure 2
8.2
Replacement of Fuses
There are two supply fuses located in the fuse drawer. To change
the fuses:
Turn power switch to the off position
Disconnect the unit from the power supply
Remove the line cord from the power entry module on the
back of the unit
Pull back on the fuse drawer catch
Pull out the fuse drawer (See Figure 3)
Check and replace with the correct fuses if necessary. The
fuses should be 5mm x 20mm quick acting, rated 250V for
230V units, 5mm x 20mm time delay, rated 250V for 115V units.
Model 281111 (115 V AC):
1.25AT
Model 281111-2 (230 V AC):
1.00AF
Push the fuse drawer back in. Reconnect unit to power
supply.