Deactivation of a pill
After usage, just go to the menu "Pill" (
) of the monitor and in the submenu
"Deactivation pill" (
Picture 24: Menu of the monitor to deactivate a
) select the capsule to stop and
press OK.
In order to ensure the success of the procedure, the capsule and monitor must be close enough
to communicate.
A confirmation has to be validated before the deactivation of the capsule. This
action is definitive; the capsule disappears from the Monitor database. The data
file corresponding to the capsule is stored until the data downloading to a computer
and activation of a new capsule on this location.
Three cases may be displayed:
"Busy slot ": The data of a capsule remains on the monitor, but the capsule is no longer
associated. Data can be downloaded on the e-Performance Manager software. As long
as unloading did not take place, a new capsule cannot be activated on this location.
"Active slot": A capsule is associated
"Free slot": This location is available to the activation of a new capsule
.
Warning: If the capsule to deactivate is not in the range of communication of the monitor at this
moment, it will not be shut down. The monitor and the capsule have to communicate to allow
this operation.
When the monitor and the pills can’t communicate during the deactivation, the monitor will
send the order during the 4 subsequent communications with the pill. Depending on the current
sampling frequency this process duration can range from 20s (Sampling rate 5s) to 20min
(Sampling rate 5min). During this period do not restart the process or else, it will extend the
waiting duration.
6.7.11
Use of the monitor screen
In order to save battery, the monitor screen will switch off automatically after 3 minutes of
inactivity when the monitor is in battery operation. This action is cancelled for 2 possible
reasons:
Picture 25: Menu of the monitor to deactivate a pill
Picture 24: Selection of a capsule to deactivate.