![BOC Edwards Seiko Seiki STPH1303C Скачать руководство пользователя страница 72](http://html1.mh-extra.com/html/boc-edwards/seiko-seiki-stph1303c/seiko-seiki-stph1303c_instruction-manual_2781984072.webp)
STP-H803/H1303 Series Instruction Manual
9 Adjustment
Methods
9.1 Tuning
For the STP pump, tuning is required to align the position of levitation in the
axial direction with the center of the rotor's movable range.
Tuning can be performed simply by pressing the "RESET" switch on the front panel.
Tuning is completed before delivery, so it is not required before use. However,
re-tuning is required in the following cases in which the position of levitation
may deviate from the center of the rotor's movable range:
◇
When changing the length of the STP connection cable (not when
changing the length of the power cable).
◇
When changing the combination of the STP pump and the STP
control unit (The error message "CAUTION: Bad EEPROM content" is
displayed upon performing the self test).
Operation
9.1.1 Tuning
Method
1) Connect each cable.
2) Switch "ON" the breaker on the rear panel.
3) After the STP pump completes the self test and enters the levitation state or
the error message "CAUTION: Bad EEPROM content" is displayed, set
MANUAL/REMOTE changeover switch to the manual and press the
"RESET" switch on the front panel for approx. 5 seconds. Tuning starts
with "Tuning!" displayed on the LCD.
When the tunable error message such as "CAUTION: Bad EEPROM content"
is displayed, it takes approx. 20 seconds until the display changes "Tuning".
4) Approx. one minute later the LCD displays "Tuning Complete", next it
displays "Levitation". This completes the tuning. (When changing the
MANUAL/REMOTE changeover switch, set to its setting before tuning.)
◇
DO NOT turn off the primary power (DO NOT switch the breaker
"OFF") during tuning. When doing so, perform tuning again.
"CAUTION
:
PCB&Pump EEPROM Failure" may be displayed after
powering ON again. When tuning is performed with this message,
the setting is initialized. Reconfigure the setting if it has been
changed by customer.
9-1