STP-H803/H1303 Series Instruction Manual
7.6
Operating the TMS Unit (For use with the TMS unit only)
7.6.1 Starting/Stopping
Power on the STP pump to automatically start the TMS unit.
Power off the STP pump to automatically stop the TMS unit.
While the TMS unit is in operation, the "TEMP CTRL" lamp of the STP control
unit lights.
When the STP pump is in the power ON state, the TMS unit operates regardless
of the status of the STP control unit. When the STP pump is the malfunction,
the TMS unit is stopped.
Operation
7.6.2 Temperature
Control
•
When the temperature of the STP pump base is lower than the setting
value, the TMS heater is turned "ON", and the cooling water value is
turned "OFF" to heat the STP pump.
•
When the temperature of the STP pump base is higher than the setting
value, the TMS heater is turned "OFF", and the cooling water value is
turned "ON" to cool the STP pump.
The LCD displays the actual temperature of the STP pump base as "TMS
TEMP:**°C."
7-7