
Remote
Local
% Speed
RPM
mA
Signal
RUN MODES
DISPLAY
M1
Remote
Start / Stop
FVS
Manual
Adjust
1. Press & Hold mA Input to Setup
2. Use Arrow Keys to Adjust
3. Press mA Input to Toggle Next
4. Press & Hold mA Input to Exit
MIN mA
MIN % Speed
MAX mA MAX % Speed
PERISTALTIC
METERING PUMP
4-20mA
Input
6.0
TOUCHPAD LAYOUT
Item Number
Item
1
LED/LCD Readout
2
Up Key
3
Rate Display Key
4
Down Key
5
Manual Adjust Control
6
4-20mA Input Control
7
Remote Start/Stop Key
8
Flow Verification Sensor (FVS) Key
9
Lock-Out Key
10
Start & Stop Key
11
Prime Key
8
7
9
11
10
3
3
2
5
6
4
1
Page 10
®
FLEXFLO M1
Page 11
Remote
Local
% Speed
RPM
mA
Signal
RUN MODES
DISPLAY
M1
Remote
Start / Stop
FVS
Manual
Adjust
1. Press & Hold mA Input to Setup
2. Use Arrow Keys to Adjust
3. Press mA Input to Toggle Next
4. Press & Hold mA Input to Exit
MIN mA
MIN % Speed
MAX mA MAX % Speed
PERISTALTIC
METERING PUMP
4-20mA
Input
5.4
Programming FVS (Flow Verification Sensor)
Item Number
Directions
1
Confirm that pump is in the OFF position
2
Press and hold FVS button
3
Toggle up/down arrow until desired
4
Press and hold FVS button to set
NOTE: Alarm Delay Time range 1-20 seconds
1
2
®
FLEXFLO M1
Содержание FLEXFLO M1 Series
Страница 1: ...Peristaltic Metering Pump Series M1...
Страница 25: ...1 2 3 4 5 6 Page 25 14 1 EXPLODED VIEW FLEXFLO M1...