Should unit require return to Blatchford, please remove and dispose of the batteries.
(Disposal as per local waste handling regulations).
11
8.4
Using the Reset Switch
The instructions in this section are for practitioner use only.
Initial Set Up, Calibration and Programming
• Selects factory default program settings
• Enters calibration mode followed by programming mode.
1.
Press reset button, confirmed by 1 beep.
2.
Within 2 seconds press reset button again, confirmed by 2 beeps.
3.
Within 2 seconds press reset button again, confirmed by 3 beeps.
Fine Tuning
• Enters program mode starting with current program settings
1.
Press reset button, confirmed by 1 beep.
2.
Within 2 seconds press reset button again, confirmed by 2 beeps.
Basic Reset
• Reverts to users programmed settings in normal operation mode.
1.
Press reset button, confirmed by 1 beep.
Note… A basic reset should always be carried out after fitting batteries.
When would you carry out any of the above resets?
• No response from the limb when trying any of the knee activation sequences
• Limb mode unknown
• After changing or reconnecting batteries (Basic Reset)
938250PK2/1-0421
Reset
Switch