69
Operations & Maintenance Manual: Product Specifications
BATTERY CHARGER/WIRELESS HUB SPECIFICATIONS
The battery charger/wireless hub is powered from a standard wall outlet (adaptable to international power
standards). The battery charger/wireless hub can charge two batteries simultaneously.
Table 27. Battery Charger/Wireless Hub Specifications
SPECIFICATION
DESCRIPTION
Operating Conditions
Ambient temperature range
5–40ºC (41–104ºF)
Atmospheric pressure range
700–1060 hPa
Relative humidity
30–75%, non‑condensing
Computer connection
USB 2.0
Charger
Powered by a desktop DC power supply.
Input voltage
18 V DC
Insulation
Class II with double insulation
Testing
IEC 60950‑1
Storage Conditions
Storage
Indoor
Ambient temperature range
‑20–60ºC (‑4–140ºF)
Atmospheric pressure range
500–1060 hPa
Relative humidity
20–95% non‑condensing
Table 28. Power Adapter Specifications
CONDITION
DESCRIPTION
Input voltage
100–240 V AC RMS
Output voltage
18 V DC
Output current
2.5 Amp
Input frequency
50–60 Hz <47–63 Hz>
Input connection
2 wire IEC 60320 C7
Fuses
250 V AC, 2 A, quick acting
Testing
IEC 60950‑1
Storage Conditions
Ambient temperature range
‑10–80ºC (14–176ºF)
Atmospheric pressure range
500–1060 hPa
Relative humidity
5–90% non‑condensing