58
PROCEdURE 3. CALIBRATE THE PROBE USING THE SCAN POINT SYSTEM
If you do not have Scan Point
®
with QuickPrint software, you must send your instrument to an authorized
Verathon
®
service center for calibration. Contact Verathon Customer Care for more information.
At minimum, the BVI 9400 must be calibrated every 12 months in order to ensure accurate results.
Calibrating ensures accurate and proper alignment of the instrument’s internal coordinate system. If
calibration is not performed by the prescribed date, the instrument can still be used to take scans but
measurements may be compromised. When calibration is required, a warning appears in the display header.
Figure 14. Calibration Warning
1. Within 3 m (10 ft) of the Battery Charger/Wireless Hub, place the calibration tank on a flat, nonreflective
surface, and then remove the lid.
2. Pour clean, room‑temperature water into the tank base, filling to the indicator mark. Ensure that there is
a minimal amount of bubbles in the water.
Note: The tank may need to sit for 24 hours until the water has degassed.
3. Using the notches to position the spiral‑shaped target correctly, place the target in the tank base.