30
To ensure reliability, a backup copy of the scan count is stored in device memory. If both the scan counter
and its backup copy are corrupted, the scan counter will automatically reset to a zero value.
The scan counter feature is designed so that the value cannot be manually reset or modified by the clinical
user or service partner.
VIEWING THE SCAN COUNTER
The scan counter can be viewed on the Self Test screen.
PRINTING THE SCAN COUNT FROM THE SELF TEST SCREEN
Once the Self Test screen is accessed, data on the screen is printed automatically using the instrument’s
onboard printer.
Figure 6. Printout of Self Test Screen