8
Figure 1.4
Terminal Block and Cable
1.3 Home screen
The home screen provides access to the various functions and settings of the unit. Access this menu via the “Home”
button in the top right corner of the screen.
Item Description
1
Setup Menu
2
Trigger Menu
3
Logic Channel Config Menu
4
XY Display
5
F(t) Display
6
Numeric Display
7
Channel Setup Menu
8
Channel and Function Display Config
9
Recording Display
10
Recording Replay Screen
11
Channel Configuration Screen
Figure 1.5
Home Screen
1.4 Recording Period
The acquisition sample rate is configurable. The unit samples one channel at a time. The sample period defines the time
a sample consumes.
Note:
Each channel is sampled in sequence. The effect is that the rate at which a channel is sampled depends on
the number of channels and the sample period of each channel. See
The sample period is set to one of the following settings: 1, 2, 5, 10, 20, 50 and 100ms. When 3-wire thermocouple
measurements (Pt100 and Pt1000) are made, the time periods are doubled. The doubled time periods are highlighted
in the channel setup dialog with a “(x2)” beside the period. Also, by using a configurable low-pass filter per channel,
measurements can be made filtering out higher frequency components.