72
calibration.
14.1.2
150V range calibration
Press ESC key to return to Setting and Measurement screen and press Configure. Select
150V for Range (V). Return to the Setting and Measurement screen, and repeat Steps 2
~ 3 to calibrate Freq1 through Freq4. Note that the HIGH calibration voltage of 260VAC
becomes 140VAC.
14.1.3
DC Voltage Calibration
Connect the DMM to the output of the instrument. Set the DMM to DCV voltage
measurement.
7.
Press System
→
Next
→
Calibration
→
softkeys, Enter the password 13579, and select
Vdc (DC voltage) on the calibration menu.
1.
Press Cal. softkey to start calibration. The instrument will output Low Point voltage.
Use the number keys to enter the voltage shown on the DMM. Use the number keys
to enter the voltage shown on the DMM. Press ENTER, and the instrument will show
“OK” after the Low Point(V).
2.
Press Cal. softkey to start calibration. The instrument will output ML Point voltage.
Use the number keys to enter the voltage shown on the DMM. Use the number keys