24
Test Frequency: 10 KHz
Range
Maximum
Display
Accuracy (DF<0.5)
Spec. Note
Inductance
DF
1000mH
999.9mH
2.0%+(Lx
/10000)%+ 8
2.0%+100/Lx+10
-
200mH
199.99mH
1.5%+(Lx
/10000)%+8
2.0%+100/Lx+10
-
20mH
19.999mH
1.5%+(Lx
/10000)%+10
3.0 %+100/Lx+15
-
2000
µ
H
1999.9
µ
H
2.0%+(Lx
/10000)%+10
8.0 %+100/Lx+20
After short cal.
Notes:
1.
1.Q Value is the reciprocal of DF.
2.
This specification is based on the measurement performed at the
test socket.
3.
DUT & Test leads shall be properly shielded to GUARD if
necessary.
4.
Lx=counts of displayed L value, e.g. L=88.88H, then Lx=8888.