background image

 

 

 
 

 

IDEAL INDUSTRIES, INC. 
TECHNICAL MANUAL 
MODELS: 61-732 

61-736 

 
 
 
The Service Information provides the following information: 

   

  Precautions and safety information 

•   Specifications     
•   Performance test procedure 
•      Calibration and calibration adjustment procedure 
•      Basic maintenance (replacing the battery) 
 

MAX

600V

COM

600V

CAT.III

V

/

 

OFF

NCV

HOLD

61-736

TRUE RMS

V

V

A

 

 

Form number:  TM61732-6 
Revision: 5. Date:    Oct 2004 

 

 

 
Form number TM61732-6 

Rev 5 October 2004 

Содержание 61-732

Страница 1: ...Precautions and safety information Specifications Performance test procedure Calibration and calibration adjustment procedure Basic maintenance replacing the battery MAX 600V COM 600V CAT III V Ω OFF NCV HOLD 61 736 TRUE RMS V V Ω A Form number TM61732 6 Revision 5 Date Oct 2004 Form number TM61732 6 Rev 5 October 2004 ...

Страница 2: ...ertifications and Compliance 2 Specifications 3 General Specification 3 Voltage Specifications 4 Current Specifications 4 Resistance Specifications 4 Continuity Specifications 4 Performance Verification 5 6 Calibration 6 7 Calibration Reference Points 8 Replacing the Batteries 9 Form number TM61732 6 Rev 5 October 2004 ...

Страница 3: ...ety Information Use the meter only as described in the Users Manual If you do not do so the protection provided by the meter may be impaired Read the Safety Information page before servicing this product In this manual a Warning identifies conditions and actions that pose hazard s to the user a Caution identifies conditions and actions that may damage the meter or the test instruments The Symbols ...

Страница 4: ... to a connector on the product that is outside the range specified for that connector Avoid electric shock To avoid injury or loss of life do not connect or disconnect probes or test leads while they are connected to a voltage source Do not operate in wet damp conditions To avoid electric shock do not operate this product in wet or damp conditions Certifications and compliances Safety Designed to ...

Страница 5: ...l alkaline 61 736 Low Battery Indicator symbol indicates low battery voltage Auto power off Approximately 10 minutes Dimensions 7 4 H X 2 6 W X 1 5 D 187mm H X 65mm W X 37mm D Maximum Cable Size ACA 1 32mm Weight Approximately 6 7 oz or 190g including battery RANGES and ACCURACY SPECIFICATION Accuracy Accuracy specifications at 23 C 5 C 73 4 F 9 F less than 75 RH Temperature Coefficient 0 1 times ...

Страница 6: ...its AC Voltage 600V 50Hz 500Hz 1 5 8 digits DC Voltage 400V 600V 0 5 2 digits 40A 400A 50Hz 60Hz 1 7 10 digits AC Current 40A 400A 60Hz 400Hz 3 0 10 digits 400Ω 4KΩ 40KΩ 400KΩ 1 0 4 digits 4MΩ 1 5 4 digits Resistance 40MΩ 3 0 5 digits Continuity 400Ω on Continuity Not specified AC Converter 61 732 is Average responding RMS Calibrated to Sine Wave 61 736 is True RMS responding Overload Protection A...

Страница 7: ...imit High Limit Model Number V AC 400V 350V AC 50Hz 345 3 354 7 61 732 V AC 400V 350V AC 50Hz 345 0 355 0 61 736 V AC 400V 350V AC 500Hz 345 3 354 7 61 732 V AC 400V 350V AC 500Hz 345 0 355 0 61 736 V AC 600V 500V AC 50Hz 487 513 61 732 V AC 600V 500V AC 50Hz 484 516 61 736 V AC 600V 500V AC 500Hz 487 513 61 732 V AC 600V 500V AC 500Hz 484 516 61 736 Table 2 DC Voltage Test Function Range Input Lo...

Страница 8: ...ion at 23 2 C 73 4 F 3 5 F at relative humidity of 70 Temperature stabilization should be reached after 30 minutes 2 Disconnect the test leads and turn the range switch to OFF 3 Remove the two screws holding the case bottom 4 Remove the case bottom using care not to damage the battery connector 5 Using a calibrated meter ensure the batteries measures a minimum of 2 4V DC If the batteries measures ...

Страница 9: ...ction range to the 400V AC 2 Connect the calibrator to the V and COM inputs on the meter 3 Output 390 0VAC 50Hz Adjust VR3 VR 200KΩ until unit display reads 3900V 1 digit 4 De energize source and remove test leads A AC Calibration 61 732 Refer to Figure 1B 61 736 Refer to Figure 2B Adjustments made under front panel label 1 Set the function range to the 400A AC 2 Set output of the AC calibrator fo...

Страница 10: ...Page 8 Figure 1A 61 732 Figure 1B 61 732 Figure 2A 61 736 Figure 2B 61 736 Form number TM61732 6 Rev 5 October 2004 ...

Страница 11: ...ps head screwdriver to remove the screws on battery cover 3 Remove batteries from the battery compartment 4 Install new 1 5V batteries R03 Size AAA Alkaline type is recommended 5 Install new batteries into compartment using care to install to proper polarity 6 Reinstall battery cover Figure 3 Form number TM61732 6 Rev 5 October 2004 ...

Отзывы: