Common Command Introduction
10
Parameters Value
Description
WVTP
<type>
Type of wave
FRQ
<frequency>
Value of frequency. If wave type is Noise or DC,
you can’t set this parameter.
PERI
<period>
Value of period. If wave type is Noise or DC, you
can’t set this parameter.
AMP
<amplitude>
Value of amplitude. If wave type is Noise or DC,
you can’t set this parameter.
OFST
<offset>
Value of offset. If wave type is Noise or DC, you
can’t set this parameter.
SYM
<symmetry>
Value of symmetry. Only when wave type is
Ramp, you can set this parameter.
DUTY
<duty>
Value of duty cycle. Only when wave type is
Square and Pulse, you can set this parameter.
PHSE
<phase>
Value of phase. If wave type is Noise or Pulse or
DC, you can’t set this parameter.
STDEV
<standard deviation >
Value of Noise wave standard deviation. Only
when wave type is Noise, you can set this
parameter.
MEAN
<mean>
Value of Noise wave mean. Only when wave type
is Noise, you can set this parameter.
IDTH
<width>
Value of width. Only when wave type is Pulse,
you can set this parameter.
RISE
<rise>
Value of rise time. Only when wave type is Pulse,
you can set this parameter.
FALL
<fall>
Value of fall time. Only when wave type is Pulse,
you can set this parameter
DLY
<delay>
Value of delay. Only when wave type is Pulse, you
can set this parameter.
HLEV
<high level>
Value of high level. If wave type is Noise or DC,
you can’t set this parameter.
LLEV
<low level>
Value of low level. If wave type is Noise or DC,
you can’t set this parameter.
BANDSTATE <switch > <bandwidth> State of noise bandwidth switch. Only when wave
type is Noise, you can set this parameter.
BANDWIDTH <bandwidth value>
Value of noise bandwidth. Only when wave type is
noise, you can set this parameter.
Query Format
<channel>:BaSic_WaVe?
<channel>={C1, C2}
Query Response
<channel>:BSWV<type>,<frequency>,<amplitude>,<offset>, <duty>,<symmetry>,
<phase>,<variance>,<mean>,<width>, <rise>, <fall>, <delay>.
Example
Change channel one wave type to ramp.
C1:BSWV WVTP, RAMP
Change frequency of channel one to 2000 Hz.
C1:BSWV FRQ, 2000
Set amplitude of channel one to 3Vpp.
C1:BSWV AMP, 3
Read channel basic wave parameters from device.
C1:BSWV?
Return: