BK Precision 2522B Скачать руководство пользователя страница 5

FEATURES

LOW COST, HIGH PERFORMANCE

B+K Precision’s Model 2522B is one of the lowest cost

digital storage oscilloscopes in the industry, yet it includes
all the basic features needed by most technicians and engi-
neers. Unlike other digital oscilloscopes, it has a full analog
mode with infinite resolution. Digital storage modes include
refresh, roll,  and single sweep. Equivalent time  storage
techniques allow storage of repetitive waveforms up  to
20 MHz. This oscilloscope is built by and backed by B+K
Precision, a company that has been selling reliable, durable,
value priced test instruments for over 50 years.

DIGITAL STORAGE FEATURES

High Resolution

2048 samples across the width of the screen (200
samples per division) assures good resolution of stored
waveforms. 1024  samples taken at sweep rates of
10

µ

s/div to 0.1

µ

s/div.

High Sampling Rate

Sampling Rate up to 10 Ms/s (Megasamples per sec-
ond) for storing one-time events down to 50 nS.

Slow Event Display

X100 time base selections extend sweep time to 50
seconds/division. Excellent for displaying slow speed
events.

Equivalent Time Sampling

Allows repetitive waveforms up to 20 MHz to be
stored.

Pretrigger Capture

Permits viewing activity before the trigger occurs.
Selectable 0%, 25%, 50%, or 75% pre-trigger view.

Plotter Output

Analog output plus TTL Pen-Down control for strip
chart recorder or X-Y plotter.

CRT FEATURES

Rectangular CRT

Rectangular CRT with large 8 x 10 centimeter viewing
area.

Convenience

Trace rotation electrically adjustable from front panel.
0%, 10%, 90%, and 100% markers for rise time meas-
urements.

DUAL TRACE FEATURES

Dual Trace

Model 2522B has two vertical input channels for dis-
playing two waveforms simultaneously. Selectable
single trace (either CH 1 or CH 2) or dual trace.
Alternate or chop sweep selectable at all sweep rates.

Sum and Difference Capability

Permits algebraic addition or subtraction of channel 1
and channel 2 waveforms, displayed as a single trace.
Useful for differential voltage and distortion measure-
ments.

HIGH FREQUENCY FEATURES

Wide Bandwidth

Conservatively-rated

3 dB bandwidth is dc to 20

MHz.

Fast Rise Time

Rise time is less than 18 nS.

Fast Sweep

Maximum sweep speed of 10 nS/div (with X10 MAG)
assures high frequencies and short-duration pulses are
displayed with high resolution.

VERTICAL FEATURES

High Sensitivity

5 mV/div sensitivity for full bandwidth. High-sen-
sitivity 1 mV/div and 2 mV/div using PULL X5 gain
control.

Calibrated Voltage Measurements

Accurate voltage measurements (

±

3%) on 10 cali-

brated ranges from 5 mV/div to 5 V/ div. Vertical gain
fully adjustable between calibrated ranges.

SWEEP FEATURES

Calibrated Time Measurements

Accurate (±3%) time measurements. The main sweep
has 21 calibrated ranges from 0.5 s/div to 0.1

µ

S/div.

Sweep time is fully adjustable between  calibrated
ranges.

X10 Sweep Magnification

Allows closer examination of waveforms, increases
maximum sweep rate to 10 nS/div.

3

Содержание 2522B

Страница 1: ...20 MHz DIGITAL STORAGE ANALOG OSCILLOSCOPE BK PRECISION INSTRUCTION MANUAL MODEL 2522B BK PRECISION...

Страница 2: ...wet 4 Use the time proven one hand in the pocket technique while handling an instrument probe Be particularly careful to avoid contacting a nearby metal object that could provide a good ground return...

Страница 3: ...further precautionary information This symbol appears in the manual where the corresponding information is given Instruction Manual for Model 2522B 20 MHz Digital Storage Analog Oscilloscope 2000 B K...

Страница 4: ...Trace Display 12 Triggering 13 Magnified Sweep Operation 15 X Y Operation 15 Video Signal Observation 15 OPERATING INSTRUCTIONS Continued Digital Storage Operation 15 Applications Guidebook 17 MAINTEN...

Страница 5: ...otter CRT FEATURES Rectangular CRT Rectangular CRT with large 8 x 10 centimeter viewing area Convenience Trace rotation electrically adjustable from front panel 0 10 90 and 100 markers for rise time m...

Страница 6: ...ng composite video waveforms TV H position can also be used as low frequency reject and TV V position can be used as high frequency reject Variable Holdoff Trigger inhibit period after end of sweep ad...

Страница 7: ...imately 500 kHz Polarity Reversal CH 2 invert Maximum Input Voltage 400 V dc ac peak HORIZONTAL AMPLIFIER Input through channel 1 input X Y mode CH 1 X axis CH 2 Y axis Sensitivity Same as vertical ch...

Страница 8: ...splay updated continually Refresh Stored data and display updated by triggered sweep Hold Freezes channel 1 and channel 2 data immediately Save CH 2 Freezes channel 2 data immediately Pretrigger Stora...

Страница 9: ...X Y used in conjunction with the X Y control and Trigger SOURCE switch to enable X Y display mode DUAL Displays the channel 1 and channel 2 signals simul taneously Dual trace mode may be either alter...

Страница 10: ...el 1 and channel 2 sweeps are chopped and displayed simultaneously normally used at slower sweep speeds When it is pushed in the two sweeps are alternately displayed one after the other normally used...

Страница 11: ...s MIN inum normal The holdoff period increases pro gressively with clockwise rotation PULL NORM TRIG When pushed in selects AUTOmatic triggering mode In this mode the oscilloscope generates sweep free...

Страница 12: ...GLE switch en gaged this switch readies the scope for single sweep operation A suitabletriggersignalarriving afterpress ing the RESET switch initiates the single sweep 35 READY Indicator Lights when t...

Страница 13: ...eform should result EQUIPMENT PROTECTION PRECAUTIONS The following precautions will help avoid damage to the oscilloscope 1 Never allow a small spot of high brilliance to remain stationary on the scre...

Страница 14: ...ace should appear on the CRT Adjust the trace brightness with the INTENSITY control and the trace sharpness with the FOCUS control SINGLE TRACE DISPLAY Either channel 1 or channel 2 may be used for si...

Страница 15: ...setting up the scope to observe a waveform it provides sweep for wave form observation until other controls can be properly set Once the controls are set operation is often switched back to the norma...

Страница 16: ...eep trigger is developed from the trigger source waveform as it crosses the threshold level in a negative going direction TIME BASE Control Set the Time Base TIME DIV control to display the desired nu...

Страница 17: ...ms div for complete frames two interlaced fields of video At most points of measurement a composite video signal is of the polarity that is the sync pulses are negative and the video is positive In th...

Страница 18: ...ingle sweep mode can be used in the pres ence of very rapidly occurring events even continuous waveforms if desired In that case the period between pushing the RE SET switch and the arrival of the tri...

Страница 19: ...ions of waveform 10 24 divisions to be more exact This output takes place at a rate that is 1 10th that of the TIME DIV setting For example if the TIME DIV setting is 10 ms div the plotter output send...

Страница 20: ...y fuseholder away from receptacle 3 When reinstalling fuseholder be sure that the fuse is installed so that the correct line voltage is selected see LINE VOLTAGE SELECTION LINE VOLTAGE SELECTION To se...

Страница 21: ...ed only as a general check of calibration accuracy not as a signal source for performing recali bration adjustments a voltage standard calibrated at several steps and of 0 3 or better accuracy is requ...

Страница 22: ...sweep speed and using X10 magnification This sweep rate is 10 ns div A rise time of less than about five divisions at this sweep speed should be calculated Calculated Measurements For observed rise t...

Страница 23: ...quency of the signal at the input jack As an example if three cycles are displayed the calculated frequency would be approximately 15 Hz This is obviously incorrect This occurs because the DSO is taki...

Страница 24: ...omers not on open account must include payment in the form of a money order or credit card For the most current flat rate repair charges contact the factory before shipping the product Return all merc...