4.

Once a waveform is digitized, it can be stored in

long-term memory by pressing the SAVE ALL or
SAVE CH2 switches. The SAVE CH2 switch imme-
diately stores the channel 2 waveform. The SAVE
ALL 
switch immediately freezes the display and
stores both waveforms. Once a waveform is stored by
engaging either of these switches, it will be stored until
the switch is disengaged or the power is turned off.

NOTES

Any of the operating modes previously dis-
cussed in analog operation (e.g. DUAL,
ADD, etc.) can be used in digital mode,
except for X–Y operation.

Time base settings of 10

µ

s/div and faster

result in a display that is acquired through
“equivalent time sampling”. This process
develops the digitized image over many
cycles of the repetitive signal; therefore,
these higher sweep speeds cannot be used
for capture of one-time events. Equivalent
time sampling is discussed in detail in Ap-
pendix II, “Unique Characteristics of Digi-
tal Storage Oscilloscopes”.

Digitizing One-Time Events

One of the most powerful features of a digital storage

oscilloscope (DSO) is its ability to capture one-time events.
To capture one-time events, single-sweep operation is em-
ployed. This is done through the use of the SINGLE switch.
When pushed, this switch releases the REFRESH or ROLL
mode if previously engaged, and readies the digital storage
circuit to receive a trigger signal — presumably the event to
be captured or some other time-related occurrence. When
the event arrives, it is stored in the memory and displayed.
The procedure is as follows:

1.

Set the oscilloscope to run in analog mode (ANA-

LOG/DIGITAL s w i t ch to ANALOG). Select
NORM triggering, and adjust the TRIG  LEVEL
control so that the unit triggers on the event to be
captured.

2.

Set the oscilloscope to digital (STORAGE) mode by

engaging the ANALOG/STORAGE switch.

3.

Initially, set the PRE TRIG switch to 50%. A more

detailed discussion of pre- and post-triggering is given
later in this section of the manual.

4.

Press the RESET switch. The READY indicator will

momentarily light as the scope awaits the arrival of the
trigger signal. When that trigger occurs, the READY
indicator will go off, and the event being monitored
will roll to the center of the display and stop.

NOTES

Depending on the sweep TIME/DIV set-
ting, the anticipated event may roll to the
center of the screen very rapidly, or very
slowly, after the READY indicator goes
out.

Single-sweep mode can be used in the pres-
ence of very rapidly occurring events, even
continuous waveforms if desired. In that
case, the period between pushing the RE-
SET switch and the arrival of the trigger
may be very short or almost instantaneous.
As a result, the READY indicator may
light for only a very short time, perhaps not
being visible at all.

Pushing the SINGLE switch automatically
disengages both the REFresh and ROLL
modes, which are continuous modes.

5.

Once the waveform is captured, it can be stored in

long-term memory through the use of the SAVE CH2
or SAVE ALL switches.

Pre- and Post-Trigger Capture

Another powerful feature of a DSO is is ability to display

“pre-trigger” information, that is, events occurring before
the arrival of a trigger event. On the Model 2522B, pre-trig-
ger operation is available in single-sweep mode by setting
the PRE TRIG switches. As shown in the above section on
“Digitizing One-Time Events”, when 50% pre-trigger op-
eration is selected, the event to be captured is situated in the
center of the memory (roughly the center of the display) after
storage. The waveform to the left of center represents activ-
ity which occurred before that trigger arrived. Pre-Trigger
selection of 25% and 75% are also available.

If the PRE-TRIG switches are released (0% pretrigger),

then no pre-trigger information is stored, and the trigger
event rolls to the extreme left of the display. In this case, all
the information displayed on the screen represents activity
after the trigger event. You may wish to use post-trigger
operation first to observe where the trigger is occurring on
the waveform of interest, and then switch to pre-trigger
mode.

Roll Mode

In this mode of operation, the waveform rolls across the

CRT from right to left (as opposed to standard oscilloscopes,
which have the trace moving from left to right) in the same
manner as most strip chart recorders. It is most commonly
used for viewing very slow events.

1.

Set up the oscilloscope in analog mode so that the event

to be observed is properly positioned on the display.
you may wish to use AUTO triggering so that the
scope continues to draw a trace even if the event is
especially slow.

2.

Switch to digital mode (ANALOG/STORAGE

switch to STORAGE), and depress the ROLL switch.
Select a TIME/DIV setting that produces a roll at the
desired speed. As the sweep speed is decreased, the
waveform will move across the screen more slowly
and the Roll feature will become more apparent. It can
also be slowed by a factor of 100 on some ranges; see
the discussion in the next section.

OPERATING INSTRUCTIONS

16

Содержание 2522B

Страница 1: ...20 MHz DIGITAL STORAGE ANALOG OSCILLOSCOPE BK PRECISION INSTRUCTION MANUAL MODEL 2522B BK PRECISION...

Страница 2: ...wet 4 Use the time proven one hand in the pocket technique while handling an instrument probe Be particularly careful to avoid contacting a nearby metal object that could provide a good ground return...

Страница 3: ...further precautionary information This symbol appears in the manual where the corresponding information is given Instruction Manual for Model 2522B 20 MHz Digital Storage Analog Oscilloscope 2000 B K...

Страница 4: ...Trace Display 12 Triggering 13 Magnified Sweep Operation 15 X Y Operation 15 Video Signal Observation 15 OPERATING INSTRUCTIONS Continued Digital Storage Operation 15 Applications Guidebook 17 MAINTEN...

Страница 5: ...otter CRT FEATURES Rectangular CRT Rectangular CRT with large 8 x 10 centimeter viewing area Convenience Trace rotation electrically adjustable from front panel 0 10 90 and 100 markers for rise time m...

Страница 6: ...ng composite video waveforms TV H position can also be used as low frequency reject and TV V position can be used as high frequency reject Variable Holdoff Trigger inhibit period after end of sweep ad...

Страница 7: ...imately 500 kHz Polarity Reversal CH 2 invert Maximum Input Voltage 400 V dc ac peak HORIZONTAL AMPLIFIER Input through channel 1 input X Y mode CH 1 X axis CH 2 Y axis Sensitivity Same as vertical ch...

Страница 8: ...splay updated continually Refresh Stored data and display updated by triggered sweep Hold Freezes channel 1 and channel 2 data immediately Save CH 2 Freezes channel 2 data immediately Pretrigger Stora...

Страница 9: ...X Y used in conjunction with the X Y control and Trigger SOURCE switch to enable X Y display mode DUAL Displays the channel 1 and channel 2 signals simul taneously Dual trace mode may be either alter...

Страница 10: ...el 1 and channel 2 sweeps are chopped and displayed simultaneously normally used at slower sweep speeds When it is pushed in the two sweeps are alternately displayed one after the other normally used...

Страница 11: ...s MIN inum normal The holdoff period increases pro gressively with clockwise rotation PULL NORM TRIG When pushed in selects AUTOmatic triggering mode In this mode the oscilloscope generates sweep free...

Страница 12: ...GLE switch en gaged this switch readies the scope for single sweep operation A suitabletriggersignalarriving afterpress ing the RESET switch initiates the single sweep 35 READY Indicator Lights when t...

Страница 13: ...eform should result EQUIPMENT PROTECTION PRECAUTIONS The following precautions will help avoid damage to the oscilloscope 1 Never allow a small spot of high brilliance to remain stationary on the scre...

Страница 14: ...ace should appear on the CRT Adjust the trace brightness with the INTENSITY control and the trace sharpness with the FOCUS control SINGLE TRACE DISPLAY Either channel 1 or channel 2 may be used for si...

Страница 15: ...setting up the scope to observe a waveform it provides sweep for wave form observation until other controls can be properly set Once the controls are set operation is often switched back to the norma...

Страница 16: ...eep trigger is developed from the trigger source waveform as it crosses the threshold level in a negative going direction TIME BASE Control Set the Time Base TIME DIV control to display the desired nu...

Страница 17: ...ms div for complete frames two interlaced fields of video At most points of measurement a composite video signal is of the polarity that is the sync pulses are negative and the video is positive In th...

Страница 18: ...ingle sweep mode can be used in the pres ence of very rapidly occurring events even continuous waveforms if desired In that case the period between pushing the RE SET switch and the arrival of the tri...

Страница 19: ...ions of waveform 10 24 divisions to be more exact This output takes place at a rate that is 1 10th that of the TIME DIV setting For example if the TIME DIV setting is 10 ms div the plotter output send...

Страница 20: ...y fuseholder away from receptacle 3 When reinstalling fuseholder be sure that the fuse is installed so that the correct line voltage is selected see LINE VOLTAGE SELECTION LINE VOLTAGE SELECTION To se...

Страница 21: ...ed only as a general check of calibration accuracy not as a signal source for performing recali bration adjustments a voltage standard calibrated at several steps and of 0 3 or better accuracy is requ...

Страница 22: ...sweep speed and using X10 magnification This sweep rate is 10 ns div A rise time of less than about five divisions at this sweep speed should be calculated Calculated Measurements For observed rise t...

Страница 23: ...quency of the signal at the input jack As an example if three cycles are displayed the calculated frequency would be approximately 15 Hz This is obviously incorrect This occurs because the DSO is taki...

Страница 24: ...omers not on open account must include payment in the form of a money order or credit card For the most current flat rate repair charges contact the factory before shipping the product Return all merc...

Страница 25: ...t be registered by completing and mailing the enclosed warranty registration card to B K Precision 1031 Segovia Circle Placentia CA 92870 within fifteen 15 days from the date of purchase Exclusions Th...

Страница 26: ...n the industry Valuable to those with little knowledge of oscilloscopes as well as the experienced technician or engineer who wishes to refresh their memory or explore new uses for oscilloscopes Downl...

Страница 27: ...2000 B K Precision Corp 481 315 9 001 Printed in Taiwan 1031 Segovia Circle Placentia CA 92870...

Отзывы: