OPERATING INSTRUCTIONS

SAFETY PRECAUTIONS

WARNING

The following precautions must be ob-
served to help prevent electric shock.

1.

When the oscilloscope is used to make measurements

in equipment that contains high voltage, there is al-
ways a certain amount of danger from electrical shock.
The person using the oscilloscope in such conditions
should be a qualified electronics technician or other-
wise trained and qualified to work in such circum-
stances. Observe the TEST INSTRUMENT SAFETY
recommendations listed on the inside front cover of
this manual.

2.

Do not operate this oscilloscope with the case removed

unless you are a qualified service technician. High
voltage up to 2100 volts is present when the unit is
operating with the case removed.

3.

The ground wire of the 3-wire ac power plug places

the chassis and housing of the oscilloscope at earth
ground. Use only a 3-wire outlet, and do not attempt
to defeat the ground wire connection or float the oscil-
loscope; to do so may pose a great safety hazard.

4.

Special precautions are required to measure or observe

line voltage waveforms with any oscilloscope. Use the
following procedure:

a. Do not connect the ground clip of the probe to

either side of the line. The clip is already at earth
ground and touching it to the hot side of the line
may “weld” or “disintegrate” the probe tip and
cause possible injury, plus possible damage to the
scope or probe.

b. Insert the probe tip into one side of the line voltage

receptacle, then the other. One side of the recepta-
cle should be “hot” and produce the waveform. The
other side of the receptacle is the ac return and no
waveform should result.

EQUIPMENT PROTECTION
PRECAUTIONS

The following precautions will help avoid
damage to the oscilloscope.

1.

Never allow a small spot of high brilliance to remain

stationary on the screen for more than a few seconds.
The screen may become permanently burned. A spot
will occur when the scope is set up for X

Y operation

and no signal is applied. Either reduce the intensity so
the spot is barely visible, apply signal, or switch back
to normal sweep operation. It is also advisable to use
low intensity with AUTO triggering and no signal
applied for long periods. A high intensity trace at the
same position could cause a line to become perma-
nently burned onto the screen.

2.

Do not obstruct the ventilating holes in the case, as this

will increase the scope’s internal temperature.

3.

Excessive voltage applied to the input jacks may dam-

age the oscilloscope. The maximum ratings of the
inputs are as follows:

CH 1 and CH 2:

400 V dc + ac peak.

EXT TRIG:

300 V dc + ac peak.

Z-AXIS INPUT:

30 V ( dc and ac peak).

4.

Always connect a cable from the ground terminal of

the oscilloscope to the chassis of the equipment under
test. Without this precaution, the entire current for the
equipment under test may be drawn through the probe
clip leads under certain circumstances. Such condi-
tions could also pose a safety hazard, which the ground
cable will prevent.

5.

The probe ground clips are at oscilloscope and earth

ground and should be connected only to the earth
ground or isolated common of the equipment under
test. To measure with respect to any point other than
the common, use CH 2 – CH 1 subtract operation
(ADD mode and invert channel 2), with the channel 2
probe to the point of measurement and the channel 1
probe to the point of reference. Use this method even
if the reference point is a dc voltage with no signal.

11

Содержание 2522B

Страница 1: ...20 MHz DIGITAL STORAGE ANALOG OSCILLOSCOPE BK PRECISION INSTRUCTION MANUAL MODEL 2522B BK PRECISION...

Страница 2: ...wet 4 Use the time proven one hand in the pocket technique while handling an instrument probe Be particularly careful to avoid contacting a nearby metal object that could provide a good ground return...

Страница 3: ...further precautionary information This symbol appears in the manual where the corresponding information is given Instruction Manual for Model 2522B 20 MHz Digital Storage Analog Oscilloscope 2000 B K...

Страница 4: ...Trace Display 12 Triggering 13 Magnified Sweep Operation 15 X Y Operation 15 Video Signal Observation 15 OPERATING INSTRUCTIONS Continued Digital Storage Operation 15 Applications Guidebook 17 MAINTEN...

Страница 5: ...otter CRT FEATURES Rectangular CRT Rectangular CRT with large 8 x 10 centimeter viewing area Convenience Trace rotation electrically adjustable from front panel 0 10 90 and 100 markers for rise time m...

Страница 6: ...ng composite video waveforms TV H position can also be used as low frequency reject and TV V position can be used as high frequency reject Variable Holdoff Trigger inhibit period after end of sweep ad...

Страница 7: ...imately 500 kHz Polarity Reversal CH 2 invert Maximum Input Voltage 400 V dc ac peak HORIZONTAL AMPLIFIER Input through channel 1 input X Y mode CH 1 X axis CH 2 Y axis Sensitivity Same as vertical ch...

Страница 8: ...splay updated continually Refresh Stored data and display updated by triggered sweep Hold Freezes channel 1 and channel 2 data immediately Save CH 2 Freezes channel 2 data immediately Pretrigger Stora...

Страница 9: ...X Y used in conjunction with the X Y control and Trigger SOURCE switch to enable X Y display mode DUAL Displays the channel 1 and channel 2 signals simul taneously Dual trace mode may be either alter...

Страница 10: ...el 1 and channel 2 sweeps are chopped and displayed simultaneously normally used at slower sweep speeds When it is pushed in the two sweeps are alternately displayed one after the other normally used...

Страница 11: ...s MIN inum normal The holdoff period increases pro gressively with clockwise rotation PULL NORM TRIG When pushed in selects AUTOmatic triggering mode In this mode the oscilloscope generates sweep free...

Страница 12: ...GLE switch en gaged this switch readies the scope for single sweep operation A suitabletriggersignalarriving afterpress ing the RESET switch initiates the single sweep 35 READY Indicator Lights when t...

Страница 13: ...eform should result EQUIPMENT PROTECTION PRECAUTIONS The following precautions will help avoid damage to the oscilloscope 1 Never allow a small spot of high brilliance to remain stationary on the scre...

Страница 14: ...ace should appear on the CRT Adjust the trace brightness with the INTENSITY control and the trace sharpness with the FOCUS control SINGLE TRACE DISPLAY Either channel 1 or channel 2 may be used for si...

Страница 15: ...setting up the scope to observe a waveform it provides sweep for wave form observation until other controls can be properly set Once the controls are set operation is often switched back to the norma...

Страница 16: ...eep trigger is developed from the trigger source waveform as it crosses the threshold level in a negative going direction TIME BASE Control Set the Time Base TIME DIV control to display the desired nu...

Страница 17: ...ms div for complete frames two interlaced fields of video At most points of measurement a composite video signal is of the polarity that is the sync pulses are negative and the video is positive In th...

Страница 18: ...ingle sweep mode can be used in the pres ence of very rapidly occurring events even continuous waveforms if desired In that case the period between pushing the RE SET switch and the arrival of the tri...

Страница 19: ...ions of waveform 10 24 divisions to be more exact This output takes place at a rate that is 1 10th that of the TIME DIV setting For example if the TIME DIV setting is 10 ms div the plotter output send...

Страница 20: ...y fuseholder away from receptacle 3 When reinstalling fuseholder be sure that the fuse is installed so that the correct line voltage is selected see LINE VOLTAGE SELECTION LINE VOLTAGE SELECTION To se...

Страница 21: ...ed only as a general check of calibration accuracy not as a signal source for performing recali bration adjustments a voltage standard calibrated at several steps and of 0 3 or better accuracy is requ...

Страница 22: ...sweep speed and using X10 magnification This sweep rate is 10 ns div A rise time of less than about five divisions at this sweep speed should be calculated Calculated Measurements For observed rise t...

Страница 23: ...quency of the signal at the input jack As an example if three cycles are displayed the calculated frequency would be approximately 15 Hz This is obviously incorrect This occurs because the DSO is taki...

Страница 24: ...omers not on open account must include payment in the form of a money order or credit card For the most current flat rate repair charges contact the factory before shipping the product Return all merc...

Страница 25: ...t be registered by completing and mailing the enclosed warranty registration card to B K Precision 1031 Segovia Circle Placentia CA 92870 within fifteen 15 days from the date of purchase Exclusions Th...

Страница 26: ...n the industry Valuable to those with little knowledge of oscilloscopes as well as the experienced technician or engineer who wishes to refresh their memory or explore new uses for oscilloscopes Downl...

Страница 27: ...2000 B K Precision Corp 481 315 9 001 Printed in Taiwan 1031 Segovia Circle Placentia CA 92870...

Отзывы: