BK Precision 2522B Скачать руководство пользователя страница 23

APPENDIX II

UNIQUE CHARACTERISTICS OF DIGITAL STORAGE OSCILLSCOPES

Digital Storage Oscilloscopes (DSO’s) use a digital sam-

pling technique to convert analog input signals to a series of
digital words that can be stored in memory. Since digital
sampling has disadvantages as well as advantages, it is
important to be aware of these unique characteristics of
DSO’s.

ALIASING

This DSO uses Real Time Sampling when sweep

TIME/DIV settings of 50 sec/div to 20

µ

s/div are selected.

Real Time Sampling simply means that samples of the input
signals are taken at equal spaces (e.g., every 0.25 ms when
the 50 ms/div range is selected). With Real Time Sampling,
a phenomena called “Aliasing” can occur when the input
signal is not sampled often enough. This causes the digitized
signal to appear to be of a lower frequency than that of the
input signal. Unless you have an idea what the input signal
is supposed to look like, you will usually be unaware that
Aliasing is occurring.

Aliasing Example

To see what actually occurs when a Digital Storage Oscil-

loscope is Aliasing, perform the following example.

1.

Apply a 10 kHz signal to the input jack and set the

sweep TIME/DIV control to 50

µ

s/div. You should see

about 5 cycles of the waveform on the display. Since
the DSO samples the input waveform 200 times per
division, each cycle is sampled 400 times.

2.

Now change the sweep TIME/DIV control to 2 ms/div.

The display should look crowded. Because the DSO
takes 200 samples per division, the sample points are
10

µ

s apart. Since the input signal is at a frequency of

10 kHz, it is being sampled 10 times per cycle. The
resulting display is too crowded to be useful, however,
it is not incorrect (it is very similar to what you would
see on a conventional analog oscilloscope).

3.

Change the setting of the sweep TIME/DIV control to

20 ms/div. Vary the frequency a slight amount (until
the display is readable) to obtain as few cycles as
possible on the CRT. If you were to calculate the
frequency of the signal from the display, you would
come up with a much lower frequency than that of the
actual frequency of the signal at the input jack. As an
example, if three cycles are displayed, the calculated
frequency would be approximately 15 Hz. This is

obviously incorrect. This occurs because the DSO is
taking one sample every 0.1 ms and a 10 kHz signal
has one cycle every 0.1 ms. What is actually happening
is that the frequency is off (not perfectly 10 kHz) by
just enough to cause the DSO to take one sample at a
slightly different place on each cycle of the waveform.

Avoiding Aliasing

Aliasing is not limited to the above example. This phe-

nomenon can occur anytime that at least two samples per
cycle are not taken (whenever the sweep TIME/DIV setting
is much too slow for the waveform being applied to the
input). Whenever the frequency of the signal is unknown,
always begin with the fastest real-time sweep speed
(20

µ

s/div) or by viewing the waveform in the analog mode

of operation first.

NOTE

Viewing one-time events or glitches is not
possible when sweep TIME/DIV settings
higher than 20

µ

s/div are selected. Viewing

one-time events poses no problem with Ali-
asing because Aliasing can occur only with
repetitive waveforms.

EQUIVALENT TIME SAMPLING

This oscilloscope uses a sampling method called Equiva-

lent Time Sampling when sweep speeds higher than
20

µ

s/div are selected. This method permits viewing of

repetitive waveforms to 20 MHz, although the maximum
digital sampling rate is 10 Msamples/sec. When the Equiva-
lent Time Sampling mode is activated, one sample is taken
during each cycle. Of course if one sample is taken during
each cycle at the trigger point (the same point on each cycle),
only a flat trace would be produced. Therefore, it is neces-
sary to take each sample further (in time) from the trigger
point than the last sample. This incremental delay is deter-
mined by the sweep TIME/DIV control setting. Because
1024 (1 k) samples are needed to fill the display, the oscil-
loscope must sample 1024 cycles of the waveform.

Therefore, only repetitive waveforms should be observed

in this mode. Irregularities that are present on an otherwise
repetitive waveform are not likely to show up when the
Equivalent Time Samplling method is used. With only one
sample being taken during each cycle, it is very likely that
glitches and other irregularities will be skipped over.

21

Содержание 2522B

Страница 1: ...20 MHz DIGITAL STORAGE ANALOG OSCILLOSCOPE BK PRECISION INSTRUCTION MANUAL MODEL 2522B BK PRECISION...

Страница 2: ...wet 4 Use the time proven one hand in the pocket technique while handling an instrument probe Be particularly careful to avoid contacting a nearby metal object that could provide a good ground return...

Страница 3: ...further precautionary information This symbol appears in the manual where the corresponding information is given Instruction Manual for Model 2522B 20 MHz Digital Storage Analog Oscilloscope 2000 B K...

Страница 4: ...Trace Display 12 Triggering 13 Magnified Sweep Operation 15 X Y Operation 15 Video Signal Observation 15 OPERATING INSTRUCTIONS Continued Digital Storage Operation 15 Applications Guidebook 17 MAINTEN...

Страница 5: ...otter CRT FEATURES Rectangular CRT Rectangular CRT with large 8 x 10 centimeter viewing area Convenience Trace rotation electrically adjustable from front panel 0 10 90 and 100 markers for rise time m...

Страница 6: ...ng composite video waveforms TV H position can also be used as low frequency reject and TV V position can be used as high frequency reject Variable Holdoff Trigger inhibit period after end of sweep ad...

Страница 7: ...imately 500 kHz Polarity Reversal CH 2 invert Maximum Input Voltage 400 V dc ac peak HORIZONTAL AMPLIFIER Input through channel 1 input X Y mode CH 1 X axis CH 2 Y axis Sensitivity Same as vertical ch...

Страница 8: ...splay updated continually Refresh Stored data and display updated by triggered sweep Hold Freezes channel 1 and channel 2 data immediately Save CH 2 Freezes channel 2 data immediately Pretrigger Stora...

Страница 9: ...X Y used in conjunction with the X Y control and Trigger SOURCE switch to enable X Y display mode DUAL Displays the channel 1 and channel 2 signals simul taneously Dual trace mode may be either alter...

Страница 10: ...el 1 and channel 2 sweeps are chopped and displayed simultaneously normally used at slower sweep speeds When it is pushed in the two sweeps are alternately displayed one after the other normally used...

Страница 11: ...s MIN inum normal The holdoff period increases pro gressively with clockwise rotation PULL NORM TRIG When pushed in selects AUTOmatic triggering mode In this mode the oscilloscope generates sweep free...

Страница 12: ...GLE switch en gaged this switch readies the scope for single sweep operation A suitabletriggersignalarriving afterpress ing the RESET switch initiates the single sweep 35 READY Indicator Lights when t...

Страница 13: ...eform should result EQUIPMENT PROTECTION PRECAUTIONS The following precautions will help avoid damage to the oscilloscope 1 Never allow a small spot of high brilliance to remain stationary on the scre...

Страница 14: ...ace should appear on the CRT Adjust the trace brightness with the INTENSITY control and the trace sharpness with the FOCUS control SINGLE TRACE DISPLAY Either channel 1 or channel 2 may be used for si...

Страница 15: ...setting up the scope to observe a waveform it provides sweep for wave form observation until other controls can be properly set Once the controls are set operation is often switched back to the norma...

Страница 16: ...eep trigger is developed from the trigger source waveform as it crosses the threshold level in a negative going direction TIME BASE Control Set the Time Base TIME DIV control to display the desired nu...

Страница 17: ...ms div for complete frames two interlaced fields of video At most points of measurement a composite video signal is of the polarity that is the sync pulses are negative and the video is positive In th...

Страница 18: ...ingle sweep mode can be used in the pres ence of very rapidly occurring events even continuous waveforms if desired In that case the period between pushing the RE SET switch and the arrival of the tri...

Страница 19: ...ions of waveform 10 24 divisions to be more exact This output takes place at a rate that is 1 10th that of the TIME DIV setting For example if the TIME DIV setting is 10 ms div the plotter output send...

Страница 20: ...y fuseholder away from receptacle 3 When reinstalling fuseholder be sure that the fuse is installed so that the correct line voltage is selected see LINE VOLTAGE SELECTION LINE VOLTAGE SELECTION To se...

Страница 21: ...ed only as a general check of calibration accuracy not as a signal source for performing recali bration adjustments a voltage standard calibrated at several steps and of 0 3 or better accuracy is requ...

Страница 22: ...sweep speed and using X10 magnification This sweep rate is 10 ns div A rise time of less than about five divisions at this sweep speed should be calculated Calculated Measurements For observed rise t...

Страница 23: ...quency of the signal at the input jack As an example if three cycles are displayed the calculated frequency would be approximately 15 Hz This is obviously incorrect This occurs because the DSO is taki...

Страница 24: ...omers not on open account must include payment in the form of a money order or credit card For the most current flat rate repair charges contact the factory before shipping the product Return all merc...

Страница 25: ...t be registered by completing and mailing the enclosed warranty registration card to B K Precision 1031 Segovia Circle Placentia CA 92870 within fifteen 15 days from the date of purchase Exclusions Th...

Страница 26: ...n the industry Valuable to those with little knowledge of oscilloscopes as well as the experienced technician or engineer who wishes to refresh their memory or explore new uses for oscilloscopes Downl...

Страница 27: ...2000 B K Precision Corp 481 315 9 001 Printed in Taiwan 1031 Segovia Circle Placentia CA 92870...

Отзывы: