Pulse Oximeter
Malfunctions
65
10 Malfunctions
WARNING! Insufficient Monitoring by Malfunctions!
With the occurrence of function failure the patient cannot be correctly monitored in
certain cases.
The cause of the function failure must be removed as soon as possible.
Assure monitoring of patient by other means.
Action at Function Failure:
1.
Secure monitoring of patient by other means.
2.
Remove unit from patient.
3.
Correct failure with help of Table below.
WARNING! Danger to the Patient!
Never use a malfunctioning unit.
List of possible Malfunctions /Error Messages:
Message
Cause
Solution
Unit cannot be turned on.
Batteries are empty.
Connect unit to electricity and
charge the battery 10 min at
least.
Should function not occur even
with a networked connection,
inform medical device supplier.
When in use buttons do not
work.
Internal failure.
Inform medical device supplier.
Error message after self-test
run and after unit is turned on.
Internal failure.
Turn the unit ON and OFF.
Inform medical device supplier.
Defective loud speaker!
Defective alarm
loudspeaker.
Inform medical device supplier.
Sensor not connected!
The patient cable is not or
incorrectly connected to
the unit.
The connection between
sensor input and patient
cable is interrupted.
Check connection between unit
and sensor. If needed, replace
sensor or patient cable.
Sensor defect!
The unit has determined
the sensor does not work
or only in limited capacity.
Exchange the sensor with a
Masimo-Sensor!
No data
The unit has determined
the sensor or patient cable
does not work properly.
Exchange the sensor or/and
patient cable with a new one!